NI PXIe-4463 Calibration Procedure
|
© National Instruments
|
7
Analog Output Performance Verification
This section verifies the analog output (AO) performance of the NI PXIe-4463.
AO Offset Accuracy Verification
Complete the following procedure to verify the AO offset accuracy.
1.
Connect channel AO 0 of the NI PXIe-4463 to the DMM using the cables recommended in
Table 1.
2.
Configure the DMM using the information in Table 2.
3.
Generate a voltage with the NI PXIe-4463:
a.
Configure an AO voltage channel task using the values in Table 3.
b.
Create an array of 512 points, and fill every element with value 0.
c.
Write the array to AO, and start the task.
4.
Measure the output voltage with the DMM.
5.
Stop the task.
Table 2.
DMM Configuration
Configuration
Value
Function
DC Volts (DCV)
Range
1 V
Number of Power Line Cycles (NPLC)
100
Table 3.
AO Setup
Configuration
Value
Physical Channels
Dev_name
/ao0
AO Gain
0 dB
Output Terminal Configuration
Differential
Rate
51200 S/s
Sample Mode
Continuous Sample
Samples Per Channel
512