NI PXIe-4463 Calibration Procedure
|
© National Instruments
|
9
AO Gain Accuracy Verification
Complete the following procedure to verify the AO gain accuracy.
1.
Connect channel AO 0 of the NI PXIe-4463 to the DMM using the cables recommended in
Table 1.
2.
Use the values in Table 6 to configure the DMM.
3.
Generate a waveform with the NI PXIe-4463.
a.
Configure an AO voltage channel task using the values in Table 7.
Table 6.
DMM Configuration
Configuration
Value
Function
AC Volts (ACV)
Range
*
10 V
RMS Mode (SETACV)
3 (SYNC)
AC Bandwidth (ACBAND)
20,100000 (20 Hz to 100 kHz)
Level Filter (LFILTER)
1 (ON)
*
Refer to Table 9 for details about the correlation between AO Gain, DMM Range, and Signal
Amplitude.
Table 7.
AO Setup
Configuration
Value
Physical Channels
Dev_name
/ao0
AO Gain
*
0 dB
Output Terminal Configuration
Differential
Rate
51200 S/s
Sample Mode
Continuous Samples
Samples Per Channel
512
*
Refer to Table 9 for details about the correlation between AO Gain, DMM Range, and Signal
Amplitude.