16
|
ni.com
|
NI PXIe-4463 Calibration Procedure
b.
Call
Basic Function Generator
using the parameters in Table 20.
c.
Write the resulting waveform to AO, and start the task.
4.
Measure the output current with the DMM.
5.
Stop the task.
6.
7.
Compare the results with the following test limit (min):
output current
70.71 mA
rms
Timebase Frequency Accuracy Verification
Note
Both analog outputs on a single device use the same timebase circuitry.
Therefore, the measurements made on one channel are valid for both channels.
Complete the following procedure to verify the timebase frequency accuracy.
1.
Connect channel AO 0 of the NI PXIe-4463 to the frequency counter using the cables
recommended in Table 1.
2.
Configure the frequency counter using the values in Table 21.
Table 20.
Basic Function Generator Parameters
Parameter
Value
Signal Type
Sine Wave
Frequency
100 Hz
Amplitude
4.5 V
Sampling Info»Sampling Rate
51200 S/s
Sampling Info»Number of Samples
512
Table 21.
Frequency Counter Setup
Configuration
Value
Function
Frequency
Impedance
1 M
Ω
Range
50 V
Coupling
AC
Bandwidth Limit
100 kHz
Trigger Source
Internal
Gate Time
0.5 s