NI PXIe-4463 Calibration Procedure
|
© National Instruments
|
5
Note
The settings listed in this calibration procedure apply to the recommended
equipment only. If different equipment is used, ensure that it is properly configured
so that requirements listed in Table 1 are met.
Test Conditions
The following setup and environmental conditions are required to ensure the NI PXIe-4463
meets published specifications.
•
Keep connections to the NI PXIe-4463 as short as possible. Long cables and wires act as
antennas, picking up extra noise that can affect measurements.
•
Verify that all connections to the NI PXIe-4463 are secure.
•
Maintain an ambient temperature of 23 °C ± 5 °C. The NI PXIe-4463 temperature will be
greater than the ambient temperature.
•
As Found Test Limits are valid for a device temperature within ±5 °C of the last
self-calibration.
•
Keep relative humidity below 80%.
•
Allow a warm-up time of at least 15 minutes to ensure that the NI PXIe-4463 circuitry is at
a stable operating temperature.
•
Allow adequate warm-up time for all of the instruments and equipment according to the
manufacturer instructions.
BNC (F) to
Banana
Adapter
(quantity 1)
Pomona
Electronics 1269
AO Offset
Verification
AO Gain Verification
AO Flatness
Verification
AO Interchannel Gain
Mismatch Verification
AO Current Drive
Verification
AO Offset and Gain
Adjustment
—
BNC F-M-F
Tee Connector
(quantity 3)
Pomona
Electronics 3285
AO Noise Verification
Characteristic
Impedance: 50
Ω
Table 1.
Required Equipment (Continued)
Equipment
Recommended
Model
Where Used
Requirements