10
|
ni.com
|
NI PXIe-5601 Calibration Procedure
•
Frequency span: 10 Hz
•
Resolution bandwidth: 200 Hz
•
Center frequency: 53 MHz, 187.5 MHz
5.
Calculate the
IF Calibration Factor
(dB) for each frequency using the following formula:
IF Calibration Factor = Channel A Power + RF Input Calibration Factor -
Spectrum Analyzer Power
Use the appropriate frequency for the channel A power sensor and the appropriate
RF Input
Calibration Factor.
As-Found and As-Left Limits
The as-found limits are the published specifications for the NI 5601. NI uses these limits to
determine whether the NI 5601 meets the device specifications when it is received for
calibration.
The as-left limits are equal to the published NI specifications for the NI 5601, less guard bands
for temperature drift and drift over time. NI uses these limits to determine whether the NI 5601
meets the device specifications over its calibration interval.
Note
Where there is no difference between an “As Found” and an “As Left” limit,
only a “Test Limit” is specified.
Verification
Verification of the NI 5601 is complete only after you have successfully completed all
procedures in this section.
Note
Limits in the following sections are based upon the August 2009 edition of
the
NI PXIe-5601 Specifications
. Refer to the most recent specifications online at
ni.com/manuals
. If a more recent edition of the specifications is available,
recalculate the limits based upon the latest specifications.
Absolute Gain Accuracy Verification
Reference Levels Less than -30 dBm
Note
Compare the results from these measurements with the specification for
reference levels <-49 dBm in Table 6, or to the specification for reference levels
≥
-49 dBm and <-30 dBm in Table 7, as appropriate.
Caution
The connectors on the DUT and test equipment are fragile. Perform the
instructions in this procedure with great care to prevent damaging any DUTs or
test equipment.
1.
Connect the RF source to the power splitter input through the 3.5 mm-to-3.5 mm cable.
2.
Connect the power splitter reference output to power sensor A.