NI PXIe-5601 Calibration Procedure
5.
Connect the other power splitter output to the NI 5601 through the 3.5 mm-to-3.5 mm
adapter. Figure 7 shows the completed equipment setup.
Figure 7.
Instantaneous Bandwidth Equipment Setup
Note
The figure shows the NI PXI-5652, but you can also use the NI PXIe-5652.
6.
Create a new session for the NI 5601 using the niRFSA Initialize With Options VI or the
niRFSA_initWithOptions
function. Set the Driver Setup portion of the
option string
parameter to
DriverSetup=Digitizer:<external>
.
7.
Lock the NI PXI/PXIe-5652 to the Reference Clock used by the external test equipment.
Use the niRFSA Configure Ref Clock VI or the
niRFSA_ConfigureRefClock
function.