NI PXIe-5601 Calibration Procedure
Test Conditions
The following setup and environmental conditions are required to ensure the NI 5601 meets
published specifications:
•
Keep connections to the device as short as possible. Long cables and wires act as antennae,
picking up extra noise that can affect measurements.
•
Verify that all connections to the device, including front panel connections, are secure.
•
Maintain an ambient temperature of 23 °C ±5 °C.
•
Keep the relative humidity between 10% and 90%, noncondensing.
Spectrum analyzer
Rohde & Schwarz FSU26
Frequency range: 10 MHz to 6.6 GHz
Power level: -60 dBm to 20 dBm
Resolution bandwidth: 10 Hz to 1 MHz
Phase noise (1 GHz, typical):
100 Hz ......................... -100 dBc/Hz
1 kHz ........................... -120 dBc/Hz
10 kHz ......................... -130 dBc/Hz
100 kHz ....................... -130 dBc/Hz
1 MHz ......................... -142 dBc/Hz
PXI Express chassis
NI PXIe-1065 or
NI PXIe-1075
—
PXI Express controller
NI PXIe-8105
—
SMA (m)-to-SMA (f)
right angle adapter
Huber + Suhner
53-SMA-50-0-2/111_N
—
3.5 mm (m)-to-3.5 mm (m)
adapter
Huber + Suhner
32_PC35-50-0-2/199_NE
—
3.5 mm (f) power splitter
(2 resistor type)
Aeroflex 1593
—
SMA (m)-to-SMA (m) cable
(36 in.) (2 required)
Huber + Suhner
ST-18/SMAm/SMAm/36
—
3.5 mm (m)-to-3.5 mm (f)
30 dB attenuator
Huber + Suhner
6630 SMA-50-1/199NE
—
3.5 mm (m) 50
Ω
terminator
(2 required)
NI part number
778353-01 (Applied
Specialties 8018-6005),
or equivalent
—
Table 1.
Required Equipment for NI 5601 Calibration (Continued)
Equipment
Recommended Model
Minimum Requirements