NI PXIe-5601 Calibration Procedure
Initial Setup
Refer to the
NI 5661/5663/5663E RF Vector Signal Analyzers Getting Started Guide
for
information about how to install the software and hardware and how to configure the device in
MAX.
The NI PXI/PXIe-5652 must be selected as the LO for the NI 5601 in MAX.
Test System Characterization
These procedures characterize the test system response. Use the results of these procedures to
complete the
.
Measure the power splitter (and attenuation) response at the RF input frequencies used in the
verification procedures. Measure the spectrum analyzer response at the IF output frequencies
used in the verification procedures.
Power Splitter Reference Output
Designate either of the two outputs of the power splitter as the reference output.
Power Sensor Zero/Calibration
1.
Connect the power meter channel A to power sensor A.
2.
Connect the power meter channel B to power sensor B.
3.
Zero and calibrate the power sensors using the built-in power meter function.
RF Input Calibration
This section describes how to calibrate the RF input that you use to verify the NI 5601.
Note
Perform the
prior to this step. You must also
define the
Power Splitter Reference Output
prior to this step.
RF Input Calibration (Without Attenuator)
Use the results of this procedure in the following procedures:
•
Absolute Gain Accuracy Verification
•
•
Instantaneous Bandwidth Verification
•
•
Reference Level Adjustment for RF Frequencies
•
Reference Level Adjustment for RF Frequencies
•
IF Attenuation Calibration Adjustment
1.
Connect the power splitter input to the RF source output through the 3.5 mm-to-3.5 mm
cable.
2.
Connect power sensor A to the previously designated reference output of the power splitter.