NI PXIe-5601 Calibration Procedure
|
© National Instruments
|
37
7.
Initialize the IF Response calibration step using the niRFSA Initialize Calibration Step VI
or the
niRFSA_InitializeCalibrationStep
function. Set the
calibration step
parameter in the VI to
IF Response Calibration
, or set the
calibrationStep
parameter in
the function to
NIRFSA_VAL_EXT_CAL_IF_RESPONSE_CALIBRATION
.
8.
Initialize the NI PXI/PXIe-5652 using the niRFSG Initialize VI or the
niRFSG_init
function.
9.
Lock the NI PXI/PXIe-5652 to the Reference Clock used by the external test equipment.
Use the niRFSG Configure Ref Clock VI or the
niRFSG_ConfigureRefClock
function.
10. Set the NI PXI/PXIe-5652 power level to 0 dBm using the niRFSG Power Level property
or the
NIRFSG_ATTR_POWER_LEVEL
attribute.
11. Set the NI PXI/PXIe-5652 frequency to 187.5 MHz using the niRFSG Frequency property
or the
NIRFSG_ATTR_FREQUENCY
attribute.
12. Commit the settings to hardware using the niRFSG Initiate VI or the
niRFSG_Initiate
function.
13. Call the niRFSA Initiate VI or the
niRFSA_Initiate
function.
14. Configure the NI 5601 RF signal path to the following corresponding values by setting the
niRFSA RF Path Selection property or the
NIRFSA_ATTR_CAL_RF_PATH_SELECTION
attribute
according to the value referenced in Table 23 or 24 that corresponds to the module
revision of your NI 5601.
•
RF band 2
•
RF band 3
Table 23.
IF Response Settings (NI 5601 Module Revisions Prior to Revision G)
IF
Filter
RF
Freq.
(MHz)
NI 5652
LO Start
Freq.
(MHz)
NI 5652
LO Stop
Freq.
(MHz)
NI 5652
LO
Freq.
Step
(MHz)
NI 5601
LO
Freq.
(MHz)
IF1
Atten.
(dB)
IF2
Atten.
(dB)
RF
Band
187.5
MHz,
Wide
1,000
1,277.5
1,097.5
1
187.5
8
3
2
187.5
MHz,
Narrow
1,000
1,227.5
1,147.5
1
187.5
8
3
2
53
MHz
200
283
223
0.5
53
14
8
3