Appendix A
Specifications
©
National Instruments Corporation
A-7
Sampling Methods
Random interleaved sampling................ 1 GS/s down to 200 MS/s
effective sample rate, repetitive
signals only
Real-time sampling ................................ Up to 100 MS/s sample rate for
transient and repetitive signals
Power Requirements
+5 VDC .................................................. 4 A
+12 VDC ................................................ 100 mA
–12 VDC ................................................ 100 mA
Physical
Dimensions............................................. 33.8 by 9.9 cm (13.3 by 3.9 in.)
I/O connectors
Analog input CH0 ........................... BNC female
Digital triggers ................................ SMB female, 9-pin mini DIN
Operating Environment
Note
Multiple NI 5911 devices in the same computer may raise operating temperatures
beyond specification and give rise to imprecise data. NI strongly recommends leaving an
empty PCI slot between multiple NI 5911 devices or adding a fan.
Ambient temperature.............................. 5 °C to 40 °C
Relative humidity ................................... 10% to 90%, noncondensing
Storage Environment
Ambient temperature.............................. –20 °C to 65 °C