© National Instruments
|
1-11
Figure 1-9.
USB-62
xx
BNC Fuse Location
3.
Remove both end pieces by unscrewing the four sockethead cap screws with a 7/64 in. hex
wrench.
Note
The end pieces are attached using self-threading screws. Repeated screwing
and unscrewing of self-threading screws will produce a compromised connection.
4.
With a Phillips #2 screwdriver, remove the Phillips 4-40 screw adjacent to the USB
connector.
5.
Remove the nut from the power connector.
6.
Remove the four Phillips 4-40 screws that attach the top panel to the enclosure and remove
the panel and connector unit.
7.
Replace the broken fuse while referring to Figure 1-9 for the fuse location.
8.
Replace the top panel, screws, nut, and end pieces.
1
T 2A 250V (5 × 20 mm) Fuse
1
Summary of Contents for PCI-6281