NI PXI-4070 Calibration Procedure
12
ni.com
10. Set the input resistance of the NI 4070 to >10 G
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_GREATER_THAN_10_GIGAOHM
11. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 11.
12. Set the input resistance of the NI 4070 to 10 M
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_10_MEGAOHM
13. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 11.
14. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_VOLTS
•
Range
=
100
•
Resolution
=
1e–6
15. Set the input resistance of the NI 4070 to 10 M
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_10_MEGAOHM
16. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 11.
17. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_DC_VOLTS
•
Range
=
300
•
Resolution
=
300e–6
18. Set the input resistance of the NI 4070 to 10 M
Ω
by calling
niDMM_SetAttributeViReal64
with the following parameters:
•
Attribute_ID
=
NIDMM_ATTR_INPUT_RESISTANCE
•
Attribute_Value
=
NIDMM_VAL_10_MEGAOHM
19. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 11.
20. Remove the shorting bar from the NI 4070.
21. Reset the calibrator.