NI PXI-4070 Calibration Procedure
20
ni.com
10. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_VOLTS_DCCOUPLED
•
Range
=
0.05
•
Resolution
=
50e–9
11. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 12.
12. Refer to Table 2 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed under Calibrator Output
in Table 2 for the current iteration.
b.
Call
niDMM_ConfigureMeasurement
with
Mode
set to
NIDMM_VAL_AC_VOLTS
and the remaining parameters as shown
in Table 2 for the current iteration.
c.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 12.
d.
Call
niDMM_ConfigureMeasurement
again, changing
Mode
to
NIDMM_VAL_AC_VOLTS_DCCOUPLED
.
e.
Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 12.
13. Repeat step 12 for each of the remaining iterations shown in Table 2.
Table 2.
niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output
niDMM_ConfigureMeasurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
50 mV
50 Hz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
50 Hz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.05
50e–9
2
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.05
50e–9
3
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.5
500e–9
4
50 mV
20 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
20 kHz
NIDMM_VAL_AC_VOLTS
_DCCOUPLED
0.05
50e–9