National Instruments Corporation
53
NI PXI-4070 Calibration Procedure
9.
Reset the calibrator for safety reasons.
10. Call
niDMM_CalAdjustMisc
with
Type
set to
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the AC voltage modes of the NI 4070. Select
one of the following options:
•
If you want to perform additional adjustments, go to the
•
If you do not want to perform the current mode adjustments, go to the
section to verify your new calibration
coefficients before saving them to the EEPROM.
•
If you do not want to verify the adjustments you have just made, go to
the
Completing the Adjustment Procedures
section.
36
100 V
50
NIDMM_VAL_AC_VOLTS
300
50e3
100 V
50
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
50e3
37
100 V
100
NIDMM_VAL_AC_VOLTS
300
100e3
100 V
100
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
100e3
38
100 V
200
NIDMM_VAL_AC_VOLTS
300
200e3
100 V
200
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
200e3
39
50 V
300
NIDMM_VAL_AC_VOLTS
300
300e3
50 V
300
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
300e3
40
10 V
500
NIDMM_VAL_AC_VOLTS
300
500e3
10 V
500
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
500e3
Table 10.
niDMM_CalAdjustACFilter Parameters (Continued)
Iteration
Calibrator Output
niDMM_CalAdjustACFilter Parameters
Amplitude
Frequency
(kHz)
Mode
Range
(V)
Frequency
(Hz)