Table 1. PXI/PCI-5154 Front Panel Signal Connections (Continued)
Connector
Description
Function
PFI 0
Standard SMB jack
connector
Multipurpose PFI line for Reference Clock In,
Sample Clock In, and Digital Trigger In/Out.
PFI 1
Standard SMB jack
connector
Multipurpose PFI line for Reference Clock Out,
probe compensation, and Digital Trigger In/Out.
Table 2. PXI-5154 Front Panel Indicators
Label
Function
ACCESS Indicates the status of the PCI bus and the interface from the PXI-5154 to the
controller.
ACTIVE Indicates the status of the onboard acquisition hardware of the PXI-5154.
Configuring the PXI/PCI-5154 in MAX
Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which NI hardware products are in the system and how they are
configured. MAX is automatically installed with NI-SCOPE.
1.
Launch MAX.
2.
In the configuration tree, expand
Devices and Interfaces
to see the list of installed NI
hardware.
Installed modules appear under the name of their associated chassis.
3.
Expand your
Chassis
tree item.
MAX lists all modules installed in the chassis. Your default names may vary.
Note
If you do not see your module listed, press <F5> to refresh the list of
installed modules. If the module is still not listed, power off the system, ensure
the module is correctly installed, and restart.
4.
Record the identifier MAX assigns to the hardware. Use this identifier when
programming the PXI/PCI-5154.
5.
Self-test the hardware by selecting the item in the configuration tree and clicking
Self-
Test
in the MAX toolbar.
The MAX self-test performs a basic verification of hardware resources.
6.
Run the test panels on the device to verify the signal.
a)
To access the test panels, right-click the device and select
Test Panels
. The
NI-SCOPE Soft Front Panel (SFP) launches automatically.
PXI/PCI-5154 Getting Started Guide
|
© National Instruments
|
9