background image

If the ACCESS LED fails to illuminate after you power on the chassis, a problem may exist
with the chassis power rails, a hardware module, or the LED.

Caution

 

Apply external signals only while the NI 5695 is powered on. Applying

external signals while the device is powered off may cause damage.

1.

Disconnect any signals from the module front panels.

2.

Power off the chassis.

3.

Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged device.

4.

Install the module in a different chassis slot from which you removed it.

5.

Power on the chassis.

6.

Verify that the device appears in MAX.

7.

Reset the device in MAX and perform a self-test.

Device Is Too Hot

NI 5695 devices have onboard sensors that monitor the device operating temperature. When
running at room temperature (25 °C), the sensor typically reads a temperature of 30 °C ± 5 °C,
depending on the module and what other devices are installed in the chassis.

Access the onboard sensor using the ni5690 Get Temperature VI or the
ni5690_getTemperature function.

Performance Issues Using an MXI Connection

When you use an MXI-3 connection to control the PXI chassis, you must run the MXI
Optimization Application prior to using the NI 5695. By default, this application runs
automatically when Windows starts. Using an MXI connection without running this
application may result in a 

time exceeded

 or 

internal software error

, an

initialization timeout, or a performance issue when using the NI 5695.
1.

Ensure the MXI software is installed. If the MXI software is not installed, refer to the
setup instructions included with the MXI kit.

After installation, you may be required to restart your computer before using the MXI
application.

2.

For MXI-3 optimization, select 

Start

»

All Programs

»

National Instruments MXI-3

»

MXI-3 Optimization

.

MXI-4 optimization is performed automatically by the MXI-4 hardware.

If you continue to have initialization or performance issues, refer to the MXI documentation
included in your MXI kit, contact NI technical support, or visit 

ni.com/support

.

12

  | 

ni.com

  | 

NI PXI-5695 Getting Started Guide

Summary of Contents for PXI-5695

Page 1: ...Contents Electromagnetic Compatibility Guidelines 2 Unpacking the Kit 2 Preparing the Environment 3 Verifying the System Requirements 3 Verifying the Kit Contents 3 Other Equipment 4 Installing the S...

Page 2: ...product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Caution To ensure the specified EMC performance operate this product o...

Page 3: ...o the product if the product is damaged in any way If the product is damaged return it to National Instruments for repair Verifying the System Requirements To use the NI 5690 instrument driver your sy...

Page 4: ...ow does not appear navigate to the drive double click it and double click autorun exe 3 Follow the instructions in the installation prompts Note Windows users may see access and security messages duri...

Page 5: ...s System Timing Slot 5 PXI Express Peripheral Slot NI 5695 modules can be placed only in PXI peripheral slots 6 Touch any metal part of the chassis to discharge static electricity 7 Ensure that the ej...

Page 6: ...antennas connected to the NI 5695 IN connectors If you are using noninsulated devices such as a noninsulated RF antenna ensure the devices are maintained in a static free environment If you are using...

Page 7: ...S ACTIVE CH 0 IN 33 dBm MAX 50 IN 27 dBm MAX 50 CH 1 50 MHz 8 GHz OUT 50 33 dBm MAX Power Reverse OUT 50 26 dBm MAX Power Reverse 1 100 N cm SMA Torque Wrench 2 Apply Force Here Until Wrench Clicks NI...

Page 8: ...Access Function CH 0 IN Input CH 0 IN and CH 0 OUT each function as a fixed attenuator with switchable main and direct paths The main path offers fixed attenuation with high input power capability Th...

Page 9: ...to the Devices book in the NI RF Signal Conditioning Devices Help for more information about using channels and paths Configuring the NI 5695 in MAX Use Measurement Automation Explorer MAX to configu...

Page 10: ...categories and allows you to easily browse and search installed examples You can see descriptions and compatible hardware models for each example or see all the examples compatible with one particular...

Page 11: ...lect Start Control Panel System and Maintenance Device Manager Windows XP Select Start Control Panel System Hardware Device Manager 5 If you are using a PXI controller verify that a National Instrumen...

Page 12: ...e onboard sensor using the ni5690 Get Temperature VI or the ni5690_getTemperature function Performance Issues Using an MXI Connection When you use an MXI 3 connection to control the PXI chassis you mu...

Page 13: ...m manuals Located using the NI Example Finder EXPLORE LEARN CREATE DISCOVER Support ni com support RF Solutions ni com rf Updates ni com updates Services ni com services RF Preamplifier NI PXI 5690 Ti...

Page 14: ...information support phone numbers email addresses and current events Refer to the NI Trademarks and Logo Guidelines at ni com trademarks for information on National Instruments trademarks Other produ...

Reviews: