background image

©

 National Instruments Corporation

3

NI TB-2709 Installation Guide

Do 

not

 substitute parts or modify equipment. Because of the danger of introducing 

additional hazards, do

 not

 install unauthorized parts or modify the device. Return the 

device to National Instruments for service and repair to ensure that its safety features 
are not compromised.

You 

must

 insulate all of your signal connections to the highest voltage with which the 

TB-2709 can come in contact.

Connections, including power signals to ground and vice versa, that exceed any of the 
maximum signal ratings on the terminal block can create a shock or fire hazard, or can 
damage any or all of the boards connected to the host computer and the terminal block. 
National Instruments is 

not 

liable for any damages or injuries resulting from incorrect 

signal connections.

Clean the module and accessories by brushing off light dust with a soft non-metallic brush. 
Remove other contaminants with a stiff non-metallic brush. The unit 

must 

be completely 

dry and free from contaminants before returning it to service.

Installation

The TB-2709 connects directly to the front panel of the PXI DAQ device, 
as shown in Figure 2.

Figure 2.  

Connecting the TB-2709 to an S Series Device

1

Terminal Block Mounting Screws

2

PXI Chassis

3

68-Pin I/O Connector on PXI Device

2

3

1

Summary of Contents for PXI Terminal Block NI TB-2709

Page 1: ...hannels Introduction The TB 2709 is a terminal block assembly consisting of nine SMB connectors and a 14 pin I O connector The terminal block assembly connects directly to the front panel of National...

Page 2: ...Get Started You need the following to set up and use your terminal block NI TB 2709 Terminal Block Assembly NI TB 2709 Installation Guide One of the following NI PXI 6122 NI PXI 6123 NI PXI 6132 NI PX...

Page 3: ...l ratings on the terminal block can create a shock or fire hazard or can damage any or all of the boards connected to the host computer and the terminal block National Instruments is not liable for an...

Page 4: ...Getting Started Guide for instructions on installing your PXI device 2 Guide the terminal block onto the PXI module connector Caution Do not force the terminal block when inserting it into or removing...

Page 5: ...move the switch located under the rubber sealed door on the enclosure cover for the AI channel you are using to the ON floating source switch position In the floating source switch position the ampli...

Page 6: ...the bottom of the TB 2709 provides access to digital I O signals Use the MFIT Pigtail Cable Assembly part number 194123 01 to access digital signals on the TB 2709 The MFIT Pigtail Cable Assembly fea...

Page 7: ...ssary for connecting the TB 2709 to DIO signals Figure 5 14 Pin Digital I O Connector Table 3 TB 2709 Wiring Information for 14 Pos MFIT Pigtail Cable Assembly Part Number 194123 01 Pin Number Wire Co...

Page 8: ...ages greater than 42 Vpk 60 VDC can damage the TB 2709 and any device connected to it including the host computer Overvoltage can also cause an electric shock hazard for the operator National Instrume...

Page 9: ...g peak half sine 11 ms pulse Tested in accordance with IEC 60068 2 27 Test profile developed in accordance with MIL PRF 28800F Random vibration Operating 5 to 500 Hz 0 3 grms Nonoperating 5 to 500 Hz...

Page 10: ...sources to email and phone assistance from NI Application Engineers A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manufacturer s dec...

Page 11: ...0 1 33 28 28 Malaysia 1800 887710 Mexico 01 800 010 0793 Netherlands 31 0 348 433 466 New Zealand 0800 553 322 Norway 47 0 66 90 76 60 Poland 48 22 3390150 Portugal 351 210 311 210 Russia 7 495 783 6...

Page 12: ...l Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the ap...

Reviews: