NI TB-2709 Installation Guide

Figure 4.  

NI TB-2709 Block Diagram

For more detailed descriptions of these signals, refer to the 

S Series User 

Manual

.

Additional signal configuration information is available in Measurement & 
Automation Explorer (MAX). To access this information in MAX, select 
your device under 

Devices and Interfaces

, and click the 

Device Routes

 

tab.

Measuring Floating Signals

You can use the TB-2709 to measure floating and ground-referenced 
analog input signals. To measure floating signal sources, move the switch 
located under the rubber-sealed door on the enclosure cover for the AI 
channel you are using to the ON (floating source) switch position. In the 
floating source switch position, the amplifier negative terminal connects to 
ground through a 5 k

Ω

resistor in parallel with a 0.1 

μ

F capacitor. Table 

shows the TB-2709 switch configuration options.

Table 1.  

TB-2709 Signal Descriptions

Signal Name

Reference

Direction

Description

AI <0..7>

AI GND

Input

Analog Input channels 0 through 7.

PFI 0/AI Start Trig

D GND

Input

As an input, this pin is a programmable 
function input (PFI).

Output

As an output, this pin is the AI Start 
Trigger signal.

AI <0..7> +

AI <0..7> –

Front Panel

SMB Connectors

DAQ Device

Signals

AI <0..7> GND

AI Start Trig

D GND

5 k

Ω

0.1 

μ

F

Summary of Contents for PXI Terminal Block NI TB-2709

Page 1: ...hannels Introduction The TB 2709 is a terminal block assembly consisting of nine SMB connectors and a 14 pin I O connector The terminal block assembly connects directly to the front panel of National...

Page 2: ...Get Started You need the following to set up and use your terminal block NI TB 2709 Terminal Block Assembly NI TB 2709 Installation Guide One of the following NI PXI 6122 NI PXI 6123 NI PXI 6132 NI PX...

Page 3: ...l ratings on the terminal block can create a shock or fire hazard or can damage any or all of the boards connected to the host computer and the terminal block National Instruments is not liable for an...

Page 4: ...Getting Started Guide for instructions on installing your PXI device 2 Guide the terminal block onto the PXI module connector Caution Do not force the terminal block when inserting it into or removing...

Page 5: ...move the switch located under the rubber sealed door on the enclosure cover for the AI channel you are using to the ON floating source switch position In the floating source switch position the ampli...

Page 6: ...the bottom of the TB 2709 provides access to digital I O signals Use the MFIT Pigtail Cable Assembly part number 194123 01 to access digital signals on the TB 2709 The MFIT Pigtail Cable Assembly fea...

Page 7: ...ssary for connecting the TB 2709 to DIO signals Figure 5 14 Pin Digital I O Connector Table 3 TB 2709 Wiring Information for 14 Pos MFIT Pigtail Cable Assembly Part Number 194123 01 Pin Number Wire Co...

Page 8: ...ages greater than 42 Vpk 60 VDC can damage the TB 2709 and any device connected to it including the host computer Overvoltage can also cause an electric shock hazard for the operator National Instrume...

Page 9: ...g peak half sine 11 ms pulse Tested in accordance with IEC 60068 2 27 Test profile developed in accordance with MIL PRF 28800F Random vibration Operating 5 to 500 Hz 0 3 grms Nonoperating 5 to 500 Hz...

Page 10: ...sources to email and phone assistance from NI Application Engineers A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manufacturer s dec...

Page 11: ...0 1 33 28 28 Malaysia 1800 887710 Mexico 01 800 010 0793 Netherlands 31 0 348 433 466 New Zealand 0800 553 322 Norway 47 0 66 90 76 60 Poland 48 22 3390150 Portugal 351 210 311 210 Russia 7 495 783 6...

Page 12: ...l Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the ap...