The following table lists the equipment NI recommends for the performance verification and
adjustment procedures. If the recommended equipment is not available, select a substitute
using the minimum requirements listed in the table.

Table 1. Recommended Test Equipment

Equipment

Recommended Models

Minimum Requirements

 

 

 

Frequency Reference Symmetricom 8040C

Output Frequency: 10 MHz

Accuracy: ≤ 1×10

-10

Frequency Counter

Keysight 53220A

Frequency Range: >10 MHz

Frequency Resolution: ≤0.001 Hz at
10 MHz

BNC-BNC Cable

BNC-SMA Cable

Test Conditions

Review and become familiar with the entire procedure before beginning the calibration
process.

The following setup and environmental conditions are required to ensure the PXIe-1092
chassis meets published specifications.

Keep cabling as short as possible. Long cables and wires act as antennae, picking up extra
noise that can affect measurements.

Verify that all connections, including front panel connections and screws, are secure.

Maintain an ambient temperature of 23 °C ± 5 °C.

Keep relative humidity below 80%.

Ensure that the PXIe-1092 chassis has been continuously powered for at least 24 hours
prior to calibration to allow the 10 MHz oscillator frequency to stabilize.

Ensure that the Frequency Reference has been continuously powered as specified by its
manual for its accuracy to meet the minimum requirements specified in Table 1.

2

  | 

ni.com

  | 

PXIe-1092 Calibration Procedure

Summary of Contents for PXIe-1092

Page 1: ...following software on the calibration system PXI Platform Services 18 1 or later You can download PXI Platform Services from ni com downloads PXI Platform Services supports programming the calibration...

Page 2: ...beginning the calibration process The following setup and environmental conditions are required to ensure the PXIe 1092 chassis meets published specifications Keep cabling as short as possible Long c...

Page 3: ...Counter gate time to 3 seconds 4 Measure the OCXO frequency using the Frequency Counter 5 Compare the measured frequency to the device specifications To determine whether the device under test meets i...

Page 4: ...te You can use a binary search algorithm to find the optimal oscillator control voltage The PXIe 1092 uses a 16 bit digital to analog convertor DAC to create the oscillator control voltage therefore y...

Page 5: ...h frequency increasing as the value increases Use a DAC value of 32768 which is in the middle of the valid range as a starting point a Using the same resource handle set the DacValue property to the n...

Page 6: ...er adjustment is needed Note If any test fails reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to Where to Go for Su...