Table 3. Adjustment Target Frequency Ranges (As-Left Limits)

Device

Calibration Measurement

Accuracy

Acceptable Frequency Range — As Left

Low Limit

High Limit

 

 

 

 

PXIe-1092

±1×10

-9

9,999,999.99 Hz

10,000,000.01 Hz

If the measured value is within the low-limit and high-limit range listed in Table 3, no
adjustment is needed.

Note

 

Test limits in Table 2 are based on the PXIe-1092 specifications. Refer to the

most recent PXIe-1092 specifications online at 

ni.com/manuals

.

Adjustment

Adjusting OCXO Frequency Accuracy

If the OCXO frequency accuracy is outside the specified range for the product, the device is
out of calibration.

A programmable voltage controls the OCXO frequency. By using the NI System
Configuration API to vary this voltage and precisely measure the frequency, you can find a
voltage that gives a frequency as close as possible to 10 MHz.

Note

 

You can use a binary search algorithm to find the optimal oscillator control

voltage. The PXIe-1092 uses a 16-bit digital-to-analog convertor (DAC) to create
the oscillator control voltage; therefore, you need 16 (at most) iterations of
adjustment and measurement to calibrate the 10 MHz oscillator.

Complete the following procedure to adjust the PXIe-1092 chassis OCXO frequency accuracy:
1.

Get a resource handle for the PXIe-1092 to calibrate using the System Configuration API.
a.

Call Initialize Session to set up a new session.

b.

Call Create Filter to create a new filter for searching for devices. Set the 

IsChassis

and 

IsNI

 attributes of the filter to 

TRUE

.

c.

Find the chassis resource name of the PXIe-1092 to calibrate using Measurement &
Automation Explorer (MAX) and set the 

Alias

 attribute of the filter to match the

chassis resource name.

d.

Call Find Hardware using the filter created in the previous steps. Set the 

Experts

parameter of this function to 

ni-pxi

. This function returns an array with exactly

4

  | 

ni.com

  | 

PXIe-1092 Calibration Procedure

Summary of Contents for PXIe-1092

Page 1: ...following software on the calibration system PXI Platform Services 18 1 or later You can download PXI Platform Services from ni com downloads PXI Platform Services supports programming the calibration...

Page 2: ...beginning the calibration process The following setup and environmental conditions are required to ensure the PXIe 1092 chassis meets published specifications Keep cabling as short as possible Long c...

Page 3: ...Counter gate time to 3 seconds 4 Measure the OCXO frequency using the Frequency Counter 5 Compare the measured frequency to the device specifications To determine whether the device under test meets i...

Page 4: ...te You can use a binary search algorithm to find the optimal oscillator control voltage The PXIe 1092 uses a 16 bit digital to analog convertor DAC to create the oscillator control voltage therefore y...

Page 5: ...h frequency increasing as the value increases Use a DAC value of 32768 which is in the middle of the valid range as a starting point a Using the same resource handle set the DacValue property to the n...

Page 6: ...er adjustment is needed Note If any test fails reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to Where to Go for Su...

Page 7: ...nts EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy an...