NI TB-2627 Installation Instructions
12
ni.com
where
V
TEMPOUT
is the output voltage of the temperature sensor,
where
T
(°F) and
T
(°C) are the temperature readings in degrees
Fahrenheit and degrees Celsius, respectively.
Accessories
Visit
ni.com
for more information about the following accessories.
Specifications
This section lists additional specifications for the NI TB-2627 used with the
NI PXI/PXIe-2527. All specifications are subject to change without notice.
Visit
ni.com/manuals
for the most current specifications.
Input Characteristics
All input characteristics are DC, AC
pk
, or a combination unless otherwise
specified.
Maximum switching voltage
Channel-to-channel..........................300 V
Channel-to-earth ..............................300 V, CAT I
Caution
This module is rated for Measurement Category I and intended to carry signal
voltages no greater than 300 V. This module can withstand up to 1,500 V impulse voltage.
Do
not
use this module for connections to signals or for measurements within
Categories II, III, or IV. Do
not
connect to MAINs supply circuits (for example, wall
outlets) of 115 or 230 VAC. Refer to the
Read Me First: Safety and Electromagnetic
Compatibility
document for more information about measurement categories.
Table 6.
Accessories for the NI TB-2627
Accessory
Manufacturer
Part Number
HV6-BAN4 DMM cable
National Instruments
192795-0R2
R
T
189 000
V
TEMPOUT
2.5
V
TEMPOUT
–
--------------------------------------
,
=
T
°
F
( )
T
°
C
( )
[
]
9
5
-----------------------
32
+
=
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com