CALIBRATION PROCEDURE

PXIe-4310

8-channel, 16-bit, 400 kS/s/ch, Ch-Ch Isolated Analog Input Module

This document contains the verification and adjustment procedures for the PXIe-4310 module.

Note

To maintain forced air cooling in the PXI Express system, refer to the 

Maintain Forced-Air Cooling Note to Users

.

Contents

Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
PXIe-4310 Verification and Adjustment.................................................................................. 3

Test Equipment................................................................................................................. 3

Connecting the TB-4310 (10V)................................................................................ 4

Test Conditions................................................................................................................. 5
Initial Setup....................................................................................................................... 5

Self-Calibration ........................................................................................................ 5
Checking Device Temperature Changes .................................................................. 6

Accuracy Verification....................................................................................................... 6
Adjustment........................................................................................................................ 8
Reverification ................................................................................................................... 9

Specifications............................................................................................................................ 9
Worldwide Support and Services ............................................................................................. 9

Software

Calibrating the PXIe-4310 requires the installation of NI-DAQmx on the calibration system. 
Driver support for calibrating the PXIe-4310 was first available in NI-DAQmx 17.1. For the list 
of devices supported by a specific release, refer to the 

NI-DAQmx Readme

, available on the 

version-specific download page or installation media.

You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx supports LabVIEW, 

LabWindows™/CVI™, C/C++, C#, and Visual Basic .NET. When you install NI-DAQmx, you 
only need to install support for the application software that you intend to use.

Summary of Contents for PXIe-4310

Page 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Page 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Page 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Page 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Page 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Page 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Page 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Page 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Page 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Page 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Reviews: