PXIe-4310 Calibration Procedure

3.

Initiate self-calibration using one of the following methods:

Click 

Self-Calibrate 

in the upper right corner of MAX.

Right-click 

NI PXIe-4310

 and select 

Self-Calibrate 

from the drop-down menu.

Note

Once self-calibration is completed, maintain the device temperature within 

±1 °C of the self-calibration temperature.

Note

NI recommends self-calibrating using MAX, although you can also 

self-calibrate programmatically.

Checking Device Temperature Changes

Device temperature changes (greater than ±10 °C since the previous external calibration or 
greater than ±1 °C since the previous self-calibration) can cause you to incorrectly verify and 
adjust your device. After self-calibrating your device as described in the 

Self-Calibration

 

section, complete the following steps to compare the current device temperature to the 
temperatures measured during the last self-calibration and external calibration in MAX.
1.

Launch MAX.

2.

Select 

My System»Devices and Interfaces»PXIe-4310

.

3.

Click the 

Settings

 tab. Temperature information is displayed in the Self-Calibration and 

External Calibration sections of the 

Settings

 tab.

Accuracy Verification

This section provides instructions for verifying the PXIe-4310 specifications.

Throughout the verification process, use Table 4 to determine if your device needs to be 
adjusted.

Since PXIe-4310 modules have many different ranges, you must check measurements for each 
available range. You must perform verification on all ranges of all analog input channels of 
PXIe-4310 modules in differential mode.

Note

The test limits used in this document assume a maximum temperature drift of 

±10 °C from the last external calibration, and a maximum temperature drift of ±1 °C 
from the last self-calibration. Refer to the 

Checking Device Temperature Changes

 

section for more information and instructions on reading your device temperature and 
comparing it against the device temperature during the last external calibration.

Complete the following steps to verify the analog input:

Note

The TB-4310 (600V) cannot be used for verification or adjustment of the 

PXIe-4310.

1.

Connect the output of the calibrator to the TB-4310 (10V) as shown in Figure 1.

2.

Connect the TB-4310 (10V) to the PXIe-4310.

Summary of Contents for PXIe-4310

Page 1: ...ce Temperature Changes 6 Accuracy Verification 6 Adjustment 8 Reverification 9 Specifications 9 Worldwide Support and Services 9 Software Calibrating the PXIe 4310 requires the installation of NI DAQm...

Page 2: ...nd reference information PXIe 4310 Specifications PXIe 4310 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informat...

Page 3: ...tute using the requirements listed in Table 1 Table 1 Recommended Equipment for PXIe 4310 Verification Equipment Recommended Model Requirements Calibrator Fluke 5700A If this instrument is unavailable...

Page 4: ...B 4310 10V Analog Signal Names Signal Names Reference Direction Description COM 0 7 Analog Input Isolated Ground These terminals are the reference point for differential analog input measurements AI 0...

Page 5: ...tures and temperature drift Allow a warm up time of at least 15 minutes to ensure that the PXIe 4310 measurement circuitry is at a stable operating temperature Ensure that the PXI PXI Express chassis...

Page 6: ...ettings tab Temperature information is displayed in the Self Calibration and External Calibration sections of the Settings tab Accuracy Verification This section provides instructions for verifying th...

Page 7: ...acquisition task 9 For each value in Table 4 repeat steps 3 through 8 10 Disconnect the calibrator from the device You have finished verifying the analog input specification on your device Note If ver...

Page 8: ...le referenceVoltage 7 5 Table 4 PXIe 4310 Analog Input Values Range V Test Point 24 Hour Limits 2 Year Limits Min Max Location Value V Lower Limit V Upper Limit V Lower Limit V Upper Limit V 10 10 Pos...

Page 9: ...tment verify that you have met the Test Conditions before returning your device to NI Refer to Worldwide Support and Services for assistance in returning the device to NI Specifications Refer to the P...

Page 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Reviews: