PXIe-4310 and TB-4310 (10V)/TB-4310 (600V)
|
© National Instruments
|
21
Storage Environment
Ambient temperature range .............................. -40 °C to 71 °C
(Tested in accordance with IEC 60068-2-1 and
IEC 60068-2-2. Meets MIL-PRF-28800F
Class 3 limits.)
Relative humidity range.................................... 5% to 95%, noncondensing
(Tested in accordance with IEC 60068-2-56.)
Shock and Vibration
Operating shock................................................ 30 g peak, half-sine, 11 ms pulse
(Tested in accordance with IEC 60068-2-27.
Meets MIL-PRF-28800F Class 2 limits.)
Random vibration
Operating .................................................. 5 Hz to 500 Hz, 0.3 g
rms
Non-operating........................................... 5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance with IEC 60068-2-64.
Non-operating test profile exceeds the
requirements of MIL-PRF-28800F, Class 3.)
Safety Voltages
Maximum Input Voltage
Connect only voltages that are within the following limits.
TB-4310 (10V)
AI+ to AI- ................................................. ±11 V
AI- to COM .............................................. ±11 V
TB-4310 (600V)
AI+ to AI- ................................................. 600 VDC/V
pk
, Measurement Category O;
300 V
rms
, Measurement Category II
AI- to COM .............................................. ±11 V