CALIBRATION PROCEDURE
PXIe-5113
This document contains the verification and adjustment procedures for the PXIe-5113. Refer
to
for more information about calibration solutions.
Contents
Required Software.....................................................................................................................1
Related Documentation.............................................................................................................2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................4
Password................................................................................................................................... 5
Calibration Interval................................................................................................................... 5
As-Found and As-Left Limits................................................................................................... 5
Calibration Overview................................................................................................................5
Test System Characterization....................................................................................................6
Verifying DC Accuracy...................................................................................................11
Verifying 50 Ω Bandwidth.............................................................................................. 18
Adjusting the PXIe-5113.................................................................................................25
Adjusting 1 MΩ Compensation Attenuator.................................................................... 25
Adjusting 1 MΩ DC Reference.......................................................................................26
Adjusting 50 Ω DC Reference........................................................................................ 27
Reverification..........................................................................................................................27
Updating Verification Date and Time..................................................................................... 28
Worldwide Support and Services............................................................................................ 28
Product Certifications and Declarations................................................................................. 28
Required Software
Calibrating the PXIe-5113 requires you to install the following software on the calibration
system:
•
Compatible version of the NI-SCOPE instrument driver. NI recommends you use the
latest version of NI-SCOPE for external calibration.
•
Supported application development environment (ADE): LabVIEW