Table 4. DC Accuracy Verification (Continued)
Config
Input
Impedance
(
Ω
)
Vertical
Range
(V
pk-pk
)
Vertical
Offset
(V)
Sample
Rate
(GS/s)
Test
Points
(V)
As-
Found
Limits
(mV)
As-Left
Limits (mV)
7
50
0.1 V
-4.95
1.5
-4.9975
±21.95
±10.008
8
50
0.1 V
-4.95
3.0
-4.9975
±21.95
±10.01275
9
50
0.2 V
0
1.5
0.095
±4.1
±2.1215
10
50
0.2 V
0
3.0
0.095
±4.1
±2.0455
11
50
0.2 V
0
1.5
-0.095
±4.1
±2.1215
12
50
0.2 V
0
3.0
-0.095
±4.1
±2.0455
13
50
0.2 V
4.9
1.5
4.995
±23.7
±10.9415
14
50
0.2 V
4.9
3.0
4.995
±23.7
±10.8655
15
50
0.2 V
-4.9
1.5
-4.995
±23.7
±10.9415
16
50
0.2 V
-4.9
3.0
-4.995
±23.7
±10.8655
17
50
0.4 V
0
1.5
0.19
±8
±4.119
18
50
0.4 V
0
3.0
0.19
±8
±4.043
19
50
0.4 V
0
1.5
-0.19
±8
±4.119
20
50
0.4 V
0
3.0
-0.19
±8
±4.043
21
50
0.4 V
4.8
1.5
4.99
±27.2
±12.759
22
50
0.4 V
4.8
3.0
4.99
±27.2
±12.683
23
50
0.4 V
-4.8
1.5
-4.99
±27.2
±12.759
24
50
0.4 V
-4.8
3.0
-4.99
±27.2
±12.683
25
50
1 V
0
1.5
0.475
±19.7
±9.2525
26
50
1 V
0
3.0
0.475
±19.7
±9.7275
27
50
1 V
0
1.5
-0.475
±19.7
±9.2525
28
50
1 V
0
3.0
-0.475
±19.7
±9.7275
29
50
1 V
4.5
1.5
4.975
±37.7
±17.3525
12
|
ni.com
|
PXIe-5113 Calibration Procedure