Table 4. DC Accuracy Verification (Continued)
Config
Input
Impedance
(
Ω
)
Vertical
Range
(V
pk-pk
)
Vertical
Offset
(V)
Sample
Rate
(GS/s)
Test
Points
(V)
As-
Found
Limits
(mV)
As-Left
Limits (mV)
53
50
10 V
2
1.5
5
±182.2
±98.5
54
50
10 V
2
3.0
5
±182.2
±95.5
55
50
10 V
-2
1.5
-5
±182.2
±98.5
56
50
10 V
-2
3.0
-5
±182.2
±95.5
57
1 M
0.1 V
0
1.5
0.0475
±2.15
±1.2975
58
1 M
0.1 V
0
3.0
0.0475
±2.15
±1.2975
59
1 M
0.1 V
0
1.5
-0.0475
±2.15
±1.2975
60
1 M
0.1 V
0
3.0
-0.0475
±2.15
±1.2975
61
1 M
0.1 V
5
1.5
5.0475
±22.15
±10.2975
62
1 M
0.1 V
5
3.0
5.0475
±22.15
±10.2975
63
1 M
0.1 V
-5
1.5
-5.0475
±22.15
±10.2975
64
1 M
0.1 V
-5
3.0
-5.0475
±22.15
±10.2975
65
1 M
0.2 V
0
1.5
0.095
±4.1
±2.435
66
1 M
0.2 V
0
3.0
0.095
±4.1
±2.435
67
1 M
0.2 V
0
1.5
-0.095
±4.1
±2.435
68
1 M
0.2 V
0
3.0
-0.095
±4.1
±2.435
69
1 M
0.2 V
5
1.5
5.095
±24.1
±11.435
70
1 M
0.2 V
5
3.0
5.095
±24.1
±11.435
71
1 M
0.2 V
-5
1.5
-5.095
±24.1
±11.435
72
1 M
0.2 V
-5
3.0
-5.095
±24.1
±11.435
73
1 M
0.4 V
0
1.5
0.19
±8
±4.67
74
1 M
0.4 V
0
3.0
0.19
±8
±4.67
75
1 M
0.4 V
0
1.5
-0.19
±8
±4.67
14
|
ni.com
|
PXIe-5113 Calibration Procedure