Table 1. PXIe-5113 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
Power sensor
Rohde & Schwarz
NRP6A
Test system
characterization
Verifications:
•
50 Ω
bandwidth
Power measurement:
•
Frequency range: 50 kHz
to 476 MHz
•
Power range: -27 dBm to
10 dBm
•
VSWR: <1.11
•
Absolute accuracy:
–
<0.048 dB for
50 kHz
–
<0.063 dB for
325 MHz to
476 MHz
•
Relative accuracy at
-4 dBm: <0.022 dB for
50 kHz to 476 MHz
Signal
generator
Rohde & Schwarz
SMA100A
Test system
characterization
Verifications:
•
50 Ω
bandwidth
Sine wave generation:
•
Amplitude: -22 dBm to
16 dBm
•
Frequency: 50 kHz to
476 MHz
•
Harmonics: <-30 dBc
•
Frequency accuracy:
±100 ppm
Power splitter
Aeroflex/Weinschel
1593
Test system
characterization
Verifications:
•
50 Ω
bandwidth
•
Amplitude: -22 dBm to
16 dBm
•
Frequency: 50 kHz to
476 MHz
•
VSWR: <1.25
50 Ω BNC
terminator (f)
Fairview
Microwave ST3B-F
Test system
characterization
•
Impedance: 50 Ω
•
Frequency: DC to
476 MHz
•
VSWR: <1.1
PXIe-5113 Calibration Procedure
|
© National Instruments
|
3