9.
Configure the signal generator to generate a sine waveform with the following
characteristics:
•
Frequency: the
Test Point Frequency
Power Splitter Characterization
table
•
Amplitude level: 2.0 dBm
10. Configure the power sensor to correct for the
Test Point Frequency
value using the power
sensor frequency correction function.
11. Wait 0.1 second for settling.
12. Use the power sensor to measure the power in dBm.
13. Repeat steps 9 through 12 for each configuration in the
Power Splitter Characterization
table, recording each result as
splitter output 2 power
, where each configuration has a
corresponding value.
14. Disconnect the power sensor and 50 Ω BNC terminator (f) from splitter output 2 and
splitter output 1.
15. Connect the power sensor to splitter output 1.
16. Connect the 50 Ω BNC terminator (f) to splitter output 2.
The following figure illustrates the hardware setup.
8
|