Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
•
PXIe-5110/5111/5113 Getting Started Guide
•
NI High-Speed Digitizers Help
•
PXIe-5113 Specifications
Visit
for the latest versions of these documents.
Test Equipment
Refer to the following table for a list of necessary equipment and model recommendations for
calibration of the PXIe-5113.
If you do not have the recommended equipment, select a substitute calibration standard using
the specifications listed in the minimum requirements column of the table.
Table 1. PXIe-5113 Test Equipment
Equipment
Recommended
Model
Where Used
Minimum Requirements
Oscilloscope
calibrator
Fluke 9500B/600
with Fluke 9530
Active Head
Verifications:
•
DC accuracy
Adjustment
Square wave generation:
•
Amplitude: 0.02 V
pk-pk
to
20 V
pk-pk
into 1 MΩ
symmetrical to ground
(0 V)
•
Frequency: 500 Hz
•
Aberrations: <2% of peak
for the first 500 ns
DC generation:
•
Amplitude: ±5.0 V into
50 Ω, ±118.0 V into
1 M Ω
•
Accuracy: ±(0.025% of
25 µV)
2
|
ni.com
|
PXIe-5113 Calibration Procedure