Table 1. PXIe-5113 Test Equipment (Continued)
Equipment
Recommended
Model
Where Used
Minimum Requirements
SMA (m)-to-
SMA (m) cable
—
Test system
characterization
Verifications:
•
50 Ω
bandwidth
•
Frequency: DC to
476 MHz
•
VSWR: <1.1
•
Length: ≤1 m
SMA (f)-to-
N (m) adapter
Fairview
Microwave SM4226
Test system
characterization
Verifications:
•
50 Ω
bandwidth
•
Frequency: DC to
476 MHz
•
VSWR: <1.05
BNC (f)-to-
N (f) adapter
Fairview
Microwave SM3526
Test system
characterization
Verifications:
•
50 Ω
bandwidth
•
Frequency: DC to
476 MHz
•
VSWR: <1.1
•
Impedance: 50 Ω
SMA (m)-to-
BNC (m)
adapter (×2)
Fairview
Microwave SM4716
Test system
characterization
Verifications:
•
50 Ω
bandwidth
•
Frequency: DC to
476 MHz
•
VSWR: <1.1
•
Impedance: 50 Ω
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-5113 meets
published specifications:
•
Allow all test instruments a warm-up time of at least the amount of time stated in their
specifications documents.
•
Allow a warm-up time of at least 15 minutes after the chassis is powered on and
NI-SCOPE is loaded and recognizes the PXIe-5113. The warm-up time ensures that the
PXIe-5113 and test instrumentation are at a stable operating temperature.
•
Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that
can affect measurements.
4
|
ni.com
|
PXIe-5113 Calibration Procedure