Table 4. Connectors
Signal
Connector Type
Description
CH 0 and CH 1 BNC female
Analog input connection; digitizes data and triggers
acquisitions.
CLK IN
SMB
Imports an external reference clock or sample clock to
the oscilloscope.
PFI 0
SMB
PFI line for digital trigger input/output, probe
compensation.
AUX 0
MHDMR
Reference clock input, reference clock output,
bidirectional digital PFI, and 3.3 V power output.
Table 5. Access LED
Color
Indications
Off
The module is not yet functional or the module has detected a problem with a power
rail.
Amber The module is being accessed.
Green The module is ready to be programmed.
Table 6. Active LED
Color
Indications
Off
Module is not armed, triggered, or acquiring a waveform.
Amber The module is armed and waiting for a Reference (Stop) Trigger.
Green The module has received a Reference (Stop) Trigger. Also indicates that the module
is acquiring a waveform.
Red
The module has detected an error. The LED remains red until the error condition is
removed. The following scenarios are examples of potential errors:
•
Detected an unlocked condition on a previously locked PLL. A PLL that is
unlocked while in reset does not show an error.
•
Unable to detect the external sample clock.
•
Detected an overheating or overpowering error.
Note
Certain driver interactions may cause the Active LED to flash red.
An error condition does not exist unless the Active LED remains red.
PXIe-5164 Getting Started Guide
|
© National Instruments
|
11