The MAX self-test performs a basic verification of hardware resources.

Related Information

What Should I Do if the PXIe-5164 Doesn't Appear in MAX?

 on page 21

PXIe-5164 Self-Calibration

For the PXIe-5164, self-calibration yields the following benefits:

Characterizes onboard sample clock phase adjustment circuitry

Characterizes onboard intermodule synchronization measurement circuitry

Corrects for DC gain and offset errors within the oscilloscope for all ranges and both
input impedance paths based on comparison to a precision, high-stability internal voltage
reference

Calibrates offset error of interleaving ADCs

Takes approximately 1 minute to complete

Running Self-Calibration

Running self-calibration enables the PXIe-5164 to yield full performance over its operating
temperature range and two-year external calibration cycle.

Before running self-calibration, warm up the PXIe-5164 for at least 15 minutes.

Note

 

Warm-up begins after the chassis and controller or PC is powered, the

module is recognized by the host, and the module is configured using the instrument
design libraries or NI-SCOPE.

Tip

 

In some RIO applications, the power consumed by the module can be

significantly higher than the default image for the module. In these cases, you can
improve performance by loading your image and configuring the PXIe-5164 before
warm-up time begins.

PXIe-5164 modules are externally calibrated at the factory, but for optimal performance, use
self-calibration in any of the following situations:

After first installing the PXIe-5164 in your chassis

After any module in the chassis is installed, uninstalled, or moved

When the system is in an environment where the ambient temperature varies or the
module temperature has drifted more than ±5 °C from the previous self-calibration

To periodically adjust for small performance drifts that occur with product aging

When 90 days have elapsed since the previous self-calibration

Tip

 

You can use the nodes on the LabVIEW Instrument Design Libraries for

Reconfigurable Oscilloscopes External Calibration palette to programmatically
return the date and module temperature of the previous self-calibration. You can also
view this information by selecting the device in MAX.

16

  | 

ni.com

  | 

PXIe-5164 Getting Started Guide

Summary of Contents for PXIe-5164

Page 1: ...ng the Environment 4 Choosing the Software 4 Software Options 4 Comparison of Software Features 5 Installing the Software 6 Installing the PXIe 5164 7 PXIe 5164 Front Panel and Pinout 10 PXIe 5164 SCB...

Page 2: ...Ie 5164 2 PXIe 5164 Safety Environmental and Regulatory Information 3 PXIe 5164 Getting Started Guide this document 4 Maintain Forced Air Cooling Note to Users Notice To prevent electrostatic discharg...

Page 3: ...783959 01 SHH19 MH19 AUX shielded single ended cable 1 m NI part number 784091 01 Mini HDMI Breakout Cable to 6 BNC 6 in NI part number 786979 01 Oscilloscope probe compensator SMB NI part number 7869...

Page 4: ...Ensure that the environment in which you are using the PXIe 5164 meets the following specifications Temperature Operating 0 C to 50 C Storage 40 C to 71 C Humidity Operating 10 to 90 noncondensing St...

Page 5: ...can use a combination of these software options for your application However when using the NI SCOPE driver and the instrument design libraries in a single application only one option can access the d...

Page 6: ...Refer to the LabVIEW Upgrade Notes for additional information about upgrading to the most recent version of LabVIEW for Windows 3 Optional Install NI LabVIEW FPGA Module Refer to the LabVIEW FPGA Modu...

Page 7: ...th one backplane connector The module may be installed into any PXI Express compatible slot 1 Ensure the AC power source is connected to the chassis before installing the module The AC power cord grou...

Page 8: ...ing slots 8 Touch any metal part of the chassis to discharge static electricity 9 Ensure that the ejector handle is in the downward unlatched position 10 Place the module edges into the module guides...

Page 9: ...ity and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 13 Cover all empty slots using EMC filler panels or fill using sl...

Page 10: ...anel and Pinout Front Panel PXIe 5164 400 MHz Oscilloscope ACCESS ACTIVE CH 0 CAT II TO 50 5 V Peak MAX 1 M 250 V Peak MAX CH 1 250 V RMS MAX INPUT 5 V MAX CLK IN PFI 0 AUX 0 10 ni com PXIe 5164 Getti...

Page 11: ...Active LED Color Indications Off Module is not armed triggered or acquiring a waveform Amber The module is armed and waiting for a Reference Stop Trigger Green The module has received a Reference Stop...

Page 12: ...nd reference for signals 4 GND Ground reference for signals 5 CLK OUT Used to export the reference clock 6 GND Ground reference for signals 7 GND Ground reference for signals 8 AUX 0 PFI 0 Bidirection...

Page 13: ...nals to the AUX 0 connector on the PXIe 5164 front panel Refer to the following figure and table for information about the SCB 19 signals when connected to the AUX 0 front panel connector 1 2 3 4 5 6...

Page 14: ...power 200 mA maximum 14 to 26 GND Ground reference for signals PXIe 5164 Mini HDMI Breakout Cable to 6 BNC Pinout You can use the Mini HDMI Breakout Cable to 6 BNC to connect digital signals to the AU...

Page 15: ...odules appear under the name of their associated chassis 3 Expand your Chassis tree item MAX lists all modules installed in the chassis Your default names may vary Note PXIe 5164 modules appear as NI...

Page 16: ...e module is configured using the instrument design libraries or NI SCOPE Tip In some RIO applications the power consumed by the module can be significantly higher than the default image for the module...

Page 17: ...nstruments 3 In the instrument header menu in the upper right corner of the panel add the PXIe 5164 to the large panel with Add Remove Devices 4 Click Auto to automatically configure device settings f...

Page 18: ...n Description 32 bit LabVIEW C Program Files x86 National Instruments LabVIEW year ProjectTemplates Source Digtzr Stream To Host NI 5164 documentation 64 bit LabVIEW C Program Files National Instrumen...

Page 19: ...to the PFI connector of the oscilloscope that generates the probe compensation signal 2 Attach the BNC adapter to the tip of the probe 3 Connect the probe tip via the BNC adapter to the BNC end of the...

Page 20: ...sed to compensate it and Compensate your probes frequently Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical support or visit ni com support Why...

Page 21: ...rt the system 4 Navigate to the Device Manager Operating System Description Windows 10 8 1 Right click the Start button and select Device Manager Windows 7 Select Start Control Panel Device Manager 5...

Page 22: ...review device specifications custom applications with an application programming interface API NI Reconfigurable Oscilloscopes Help PXIe 5164 Specifications Instrument Design Libraries for Reconfigur...

Page 23: ...ct registration facilitates technical support and ensures that you receive important information updates from NI NI corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759...

Page 24: ...EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and h...

Reviews: