2.
Power off the chassis.
3.
Remove the module from the chassis and inspect it for damage. Do not reinstall a
damaged module.
4.
Install the module in a different chassis slot from which you removed it.
5.
Power on the chassis.
Note
If you are using a PC with a device for PXI remote control system,
power on the chassis before powering on the computer.
6.
Verify that the module appears in MAX.
7.
Reset the module in MAX and perform a self-test.
What Should I Do if the PXIe-5164 Doesn't Appear in
MAX?
1.
In the MAX configuration tree, expand
Devices and Interfaces
.
2.
Expand the
Chassis
tree to see the list of installed hardware, and press <F5> to refresh
the list.
3.
If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.
4.
Navigate to the Device Manager.
Operating System
Description
Windows 10/8.1
Right-click the Start button, and select
Device Manager
.
Windows 7
Select
Start
»
Control Panel
»
Device Manager
.
5.
Verify the PXIe-5164 appears in the Device Manager.
a)
Under an NI entry, confirm that a PXIe-5164 entry appears.
Note
If you are using a PC with a device for PXI remote control system,
under
System Devices
, also confirm that no error conditions appear for the
PCI-to-PCI Bridge
.
b)
If error conditions appear, reinstall LabVIEW Instrument Design Libraries for
Reconfigurable Oscilloscopes and the PXIe-5164.
What Should I Do if the PXIe-5164 Fails the Self-Test
or Self-Calibration?
1.
Restart the system.
2.
Launch MAX, and perform the self-test or self-calibration again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
5.
Power on the chassis.
6.
Perform the self-test or self-calibration again.
PXIe-5164 Getting Started Guide
|
© National Instruments
|