5.
Verify the PXIe-5172 appears in the Device Manager.
a)
Under an NI entry, confirm that a PXIe-5172 entry appears.
Note
If you are using a PC with a device for PXI remote control system,
under
System Devices
, also confirm that no error conditions appear for the
PCI-to-PCI Bridge
.
b)
If error conditions appear, reinstall NI LabVIEW Instrument Design Libraries for
Reconfigurable Oscilloscopes and the PXIe-5172.
What Should I Do if the PXIe-5172 Fails the Self-Test
or Self-Calibration?
1.
Restart the system.
2.
Launch MAX, and perform the self-test or self-calibration again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
5.
Power on the chassis.
6.
Perform the self-test or self-calibration again.
Where To Go Next
Refer to the following resources and information that you may need as you create your own
application.
EXPLORE
LEARN
CREATE
DISCOVER
custom applications within
an application programming
interface (API).
Instrument Design Libraries
Reconfigurable Oscilloscope
Sample Projects*
about hardware features
or review device
specifications.
more about your products through ni.com.
the application
development environment (ADE)
for your application.
LabVIEW 2016 FPGA
Module Help
Getting Started with
LabVIEW
NI Reconfigurable
Oscilloscopes Help
NI Oscilloscopes
ni.com/oscilloscopes
Services
ni.com/services
Updates
ni.com/updates
Located online at ni.com/manuals
*This item is also installed with the driver software.
NI PXIe-5170R
14-Bit Oscilloscope
PXIe-5172
Specifications
Support
ni.com/support
NI-SCOPE Examples*
14
|
ni.com
|
PXIe-5172 Getting Started Guide