PXIe-5185/5186 Calibration Procedure
|
© National Instruments
|
19
2.
Configure the PXIe-5185/5186 with the following settings:
•
Vertical coupling
: DC
•
Input impedance
: 50
Ω
•
Maximum input frequency
: 3 GHz (PXIe-5185), 5 GHz (PXIe-5186)
•
Range
: 0.11 V
pk-pk
•
Sample rate
: The
Sample Rate
value from Table 6 for the current iteration
•
Minimum number of samples
: 1,048,567
3.
Configure the signal generator to generate a sine waveform with the following
characteristics:
•
Frequency
: the
Test Point
frequency value from Table 6
•
Amplitude level
: the
Test Point
amplitude value from Table 6
4.
Configure the power sensor to correct for the
Test Point
frequency using the power sensor
frequency correction function.
5.
Use the power sensor to measure the power in dBm. Record the result as
measured input
power
.
6.
Calculate the
corrected input power
using the following equation:
corrected input power
=
measured input power
+
splitter imbalance
Note
Select the
splitter imbalance
value from the list of test points from the
Test
System Characterization
section for the current
Test Point
frequency from Table 6.
7.
Use the PXIe-5185/5186 to acquire and measure the power using the Extract Single Tone
Information VI, converting the result from V
pk
to dBm. Record the result as
device input
power
.
8.
If the
Test Point
frequency from Table 6 is 50 kHz, proceed to the following step.
Otherwise, go to step 12.
9.
Calculate the
power reference
using the following equation:
power reference = device input power - corrected input power
10. Compare the power reference to the test limit for Iteration 1 in Table 6 to verify the 50
Ω
AC amplitude accuracy.
11. Go to step 14. The power error is not calculated for this configuration.
12. Calculate the
power error
using the following equation:
power error
=
device input power
-
corrected input power
-
power reference
13. Compare the power error to the test limit for Iteration 2 in Table 6 to verify the 50
Ω
analog
bandwidth.
14. Repeat steps 2 through 13 for each
Iteration
in Table 6.
15. Connect splitter output 2 of the power sensor assembly to the channel 1 50
Ω
input of the
PXIe-5185/5186 and repeat steps 2 through 13 for each
Iteration
listed in Table 6.