2
|
ni.com
|
PXIe-5185/5186 Calibration Procedure
50
Ω
and 1 M
Ω
Devices
Early versions of the PXIe-5185/5186 support only 50
Ω
input impedance. Check the part
number of your device to see what input impedance is supported by your device:
•
PXIe-5185 module part numbers 199363
x
-0
z
L and PXIe-5186 module part numbers
193537
x
-0
z
L (where
x
is any letter and
z
is any number) only support 50
Ω
input
impedance. These devices require NI-SCOPE 3.8.7 or later.
•
PXIe-5185 module part numbers 152962
x
-0
z
L and PXIe-5186 module part numbers
152961
x
-0
z
L (where
x
is any letter and
z
is any number) support both 50
Ω
and 1 M
Ω
input
impedance. These devices require NI-SCOPE 3.9.6 or later.
50
Ω
devices need to be tested for only 50
Ω
input impedance because they have no 1 M
Ω
input.
Software Requirements
Calibrating the PXIe-5185/5186 requires installing the NI-SCOPE instrument driver on the
calibration system. Refer to the
50
Ω
and 1 M
Ω
Devices
section to see which driver version your
device requires. You can download the NI-SCOPE instrument driver from the NI website at
ni.com/downloads/drivers
. NI-SCOPE supports programming a self-calibration and an
external calibration in multiple application development environments (ADEs). When you
install NI-SCOPE, you need to install support for only the ADE that you intend to use.
LabVIEW support is in the
niScope.llb
file, and all calibration functions appear in the
NI-SCOPE Calibration palette. For LabWindows
™
/CVI
™
users, the NI-SCOPE function panel,
niScope.fp
, provides access to the available functions.
For the locations of files you may need to calibrate your device, refer to the
NI-SCOPE Readme
,
which is available on the NI-SCOPE media.
Related Documentation
For information about NI-SCOPE and the PXIe-5185/5186, consult the following documents:
•
NI High-Speed Digitizers Getting Started Guide
—provides instructions for installing and
configuring the PXIe-5185/5186.
•
NI High-Speed Digitizers Help
—includes detailed information about the PXIe-5185/5186
and NI-SCOPE VIs and functions and information about creating applications using
NI-SCOPE.
•
PXIe-5185 Specifications
—provides the published specification values for the PXIe-5185.
•
PXIe-5186 Specifications
—provides the published specification values for the PXIe-5186.
These documents are installed with NI-SCOPE. You can also find the latest versions of the
documentation at
ni.com/manuals
.