You can download all required software from
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
•
PXIe-5423 Getting Started Guide
•
PXIe-5423 Specifications
•
NI Signal Generators Help
•
LabVIEW Help
Visit
for the latest versions of these documents.
Test Equipment
Refer to the following table for a list of necessary test equipment, minimum requirements for
the test equipment, and model recommendations for calibration of the PXIe-5423.
If you do not have the recommended equipment, select a substitute calibration standard using
the minimum requirements specifications listed in the table.
Table 1. PXIe-5423 Test Equipment
Equipment
Recommended
Model
Where Used
Minimum Requirements
Oscilloscope
PXIe-5160 or
PXIe-5171
Verifications:
•
Timebase
accuracy
Adjustments:
•
Timebase
Frequency resolution:
0.1 Hz
The ability to measure sine
waves ≥10 MHz with
amplitudes of 2 V
pk-pk
to
5 V
pk-pk
with 50 Ω input
impedance.
SMA (m)-to-
SMA (m) cable
—
Verifications:
•
Timebase
accuracy
Adjustments:
•
Timebase
Frequency range: DC to
100 MHz
Impedance: 50 Ω
Length: <1 m
2
|
ni.com
|
PXIe-5423 Calibration Procedure