1.
Connect one end of the SMA (m)-to-SMA (m) cable to a channel of the PXIe-5423 and
the other end to a channel of the oscilloscope.
Note
Timebase accuracy verification can be performed on either channel of
the two-channel PXIe-5423.
2.
Lock the oscilloscope to the precision frequency reference using the PXI Clock.
3.
Configure the oscilloscope to acquire a 10 MHz sine wave with the following settings:
•
Vertical range
: 5 V
pk-pk
•
Sample rate
: 250 MS/s
•
Input impedance
: 50 Ω
•
Minimum number of points
: 1,048,576 samples
4.
Configure the PXIe-5423 with the following settings:
•
Channel
: Channel of the PXIe-5423 connected to the oscilloscope
•
Output impedance
: 50 Ω
•
Output mode
: Standard Function
•
Waveform type
: Sine
•
Frequency
: 10 MHz
•
Amplitude
: 2.5 V
pk-pk
•
Offset
: 0 V
•
Reference Clock source
: Onboard Reference Clock
5.
Wait 1 second for settling.
6.
Measure the peak frequency (Hz) using the oscilloscope and the Extract Single Tone
Information VI and record this as
F
Measured
.
7.
Calculate the timebase error using the following formula:
Timebase Error
(ppm) =
FMeasured − 107
10
8.
Compare the frequency error to the appropriate limit from the
Note
Timebase accuracy verification is only required on one channel of the
two-channel PXIe-5423.
Verifying DC Voltage Accuracy
Follow this procedure to verify the DC voltage accuracy of the PXIe-5423 by comparing the
voltage sourced by the PXIe-5423 to the value measured by the DMM.
Refer to the following table as you complete the following steps.
PXIe-5423 Calibration Procedure
|
© National Instruments Corporation
|
7