5.
If you are using a PXI controller, verify that a
National Instruments
entry appears in the
system device list. Reinstall NI-RFSG and the device if error conditions appear in the list.
If you are using an MXI controller, right-click
PCI-to-PCI Bridge
, and select
Properties
from the shortcut menu to verify that the bridge is enabled.
What Should I Do if the Module Fails the Self-Test?
1.
Restart the system.
2.
Launch MAX, and perform the self-test again.
3.
Power off the chassis.
4.
Reinstall the failed module in a different slot.
5.
Power on the chassis.
6.
Perform the self-test again.
What Should I Do if the Thermal Shutdown Error
Appears?
The thermal shutdown error appears when device temperatures exceed safe limits. The
NI 5672 shuts down until temperatures fall to acceptable levels and you reset the device in
MAX.
1.
Power off the chassis that contains the device.
2.
Review the
Maintain Forced-Air Cooling Note to Users
included in the NI 5672 kit and
make any necessary adjustments to ensure that the device is effectively cooled.
3.
Reset the device in MAX.
The thermal shutdown error continues to be reported until you successfully reset the
device.