PXIe-5673 Calibration Procedure

Verifying Output Intermodulation Products

Complete the following steps to verify the output intermodulation (IMD) performance of an 
PXIe-5673 using a spectrum analyzer. Complete this test for each iteration in Table 8.
1.

Connect the PXIe-5611 RF OUT front panel connector to the spectrum analyzer RF INPUT 
front panel connector through the SMA(m)-to-SMA(m) cable.

2.

Connect the PXIe-5450 CLK IN front panel connector to any rubidium frequency reference 
rear panel BNC connector through the BNC(m)-to-SMA(m) cable.

3.

Connect the PXIe-5450 CLK OUT front panel connector to the PXI-5650/5651/5652 
REF IN/OUT front panel connector.

4.

Connect any available rubidium frequency reference rear panel BNC connector to the 
spectrum analyzer REF IN rear panel connector through the BNC(m)-BNC(m) cable.

5.

Generate two single-sideband tones simultaneously at 2 MHz (

F

1

) offset from the carrier 

and 2.3 MHz (

F

2

) offset from the carrier with the following niRFSG property settings:

Frequency (Hz): 85 MHz

Power Level (dBm): 0 dBm

Generation Mode: Arb Waveform

Power Level Type: Peak Power

Reference Clock Source: ClkIn

The power of each tone should be approximately -6 dBm.

6.

Set the output power for the given measurements in Table 8.

7.

Use the spectrum analyzer to measure the mean amplitude of the first fundamental tone. 
Name this value 

P

1

. For example, when the LO is set to 85 MHz, measure the first 

fundamental signal power at 87 MHz (LO +

F

1

). Use the following spectrum analyzer 

parameter settings:

Center frequency: (Frequency in step 5) + 

F

1

Reference level: 0 dBm

Frequency span: 0 Hz

Resolution bandwidth: 500 Hz

Sweep time: 5 ms

Reference clock source: External

8.

Use the spectrum analyzer to measure the mean amplitude of the second fundamental tone. 
Name this value 

P

2

. For example, when the LO is set to 85 MHz, measure the second 

fundamental signal powers at 87.3 MHz (LO +

F

2

). Use the following spectrum analyzer 

parameter settings:

Center frequency: (Frequency in step 5) + 

F

2

Reference level: 0 dBm

Frequency span: 0 Hz

Summary of Contents for PXIe-5673

Page 1: ...aracterizing the Power Splitter 5 As Found and As Left Limits 8 Verification 8 Verifying LO Output Power Accuracy 9 Verifying Modulation Impairments 9 Verifying Modulation Bandwidth and Impairments 11...

Page 2: ...le device To calibrate the PXIe 5450 Waveform Generator PXIe 5611 IQ Modulator or PXI 5650 5651 5652 RF Analog Signal Generator individually refer to their calibration procedures at ni com manuals Cal...

Page 3: ...ise Frequency accuracy Frequency range 50 MHz to 19 8 GHz Noise floor 152 dBm Hz to 6 6 GHz SMA m to SMA m cable 36 inches Huber Suhner H S ST 18 SMAm SMAm 36 Operating frequency DC to 18 GHz Impedanc...

Page 4: ...ppropriate for the type of RF connector you are using NI recommends using a 0 565 N m 5 lb in torque wrench for SMA connectors and a 0 90 N m 8 lb in torque wrench for 3 5 mm connectors Power splitter...

Page 5: ...or each other iteration Zero and calibrate the power meter sensors before each test Initial Setup Refer to the NI RF Signal Generators Getting Started Guide for information about how to install the so...

Page 6: ...use a cable it should be as short as possible The cable should always be used with the splitter for subsequent verification procedures in this document that require a splitter 8 Connect one available...

Page 7: ...requency ranges in 5 MHz increments including endpoints PXI 5650 85 MHz to 1 3 GHz PXI 5651 85 MHz to 3 3 GHz PXI 5652 85 MHz to 6 6 GHz Store the resulting measurements as splitter i dB 1 PXI 5650 56...

Page 8: ...to determine whether the PXIe 5673 meets the device specifications when it is received for calibration The as left limits are equal to the published NI specifications for the PXIe 5673 less guard ban...

Page 9: ...increments including endpoints PXI 5650 85 MHz to 1 3 GHz PXI 5651 85 MHz to 3 3 GHz PXI 5652 85 MHz to 6 6 GHz Store the resulting measurements If the results are within the selected test limit the...

Page 10: ...olution bandwidth 500 Hz Sweep time 5 ms Reference clock source External Note Zero span mode of the spectrum analyzer shortens test time by avoiding unnecessary frequency sweeping To obtain the averag...

Page 11: ...he ISR and CSR from step 9 for each LO frequency If the results are within the selected test limit the device has passed this portion of the verification Verifying Modulation Bandwidth and Impairments...

Page 12: ...500 Hz Sweep time 5 ms Reference clock source External 7 Use the spectrum analyzer to measure the mean output power of image using the following spectrum analyzer parameter settings Center frequency F...

Page 13: ...in Table 5 If the results are within the selected test limit the device has passed this portion of the verification Table 5 Modulation Bandwidth Impairment Verification Test Limits CW Source LO Frequ...

Page 14: ...eform Generator 2 PXIe 5611 IQ Modulator 3 PXI 5650 5651 5652 RF Analog Signal Generator PXI 5652 shown 4 BNC m to SMA m Cable 5 Power Sensor 6 Rubidium Frequency Source Rear Panel 7 BNC m to BNC m Ca...

Page 15: ...n step 8 of the Characterizing the Power Splitter section must be the same port used to connect the power splitter to the power meter in step 5 of this section 5 Connect port A of the power splitter t...

Page 16: ...n 5 dB Preamplifier niRFSG Power Level property 13 Use the power meter measurements to calculate a correction for the spectrum analyzer inaccuracies and attenuator loss with the following equation Cor...

Page 17: ...ter frequency Frequency in step 8 1 MHz Reference level 30 dBm Frequency span 0 Hz Resolution bandwidth 500 Hz Sweep time 200 ms Reference clock source External Attenuation 5 dB Preamplifier On 16 Mea...

Page 18: ...arrier and 2 3 MHz F2 offset from the carrier with the following niRFSG property settings Frequency Hz 85 MHz Power Level dBm 0 dBm Generation Mode Arb Waveform Power Level Type Peak Power Reference C...

Page 19: ...litude of the second close in IMD tone at 87 6 MHz LO 2F2 F1 Name this value I2 Use the following spectrum analyzer parameter settings Center frequency Frequency in step 5 F2 Reference level 0 dBm Fre...

Page 20: ...nt panel connector 4 Connect any available rubidium frequency reference rear panel BNC connector to the spectrum analyzer REF IN rear panel connector through the BNC m to BNC m cable 5 Generate an arb...

Page 21: ...ranges in 250 MHz increments including endpoints PXI 5650 85 MHz to 1 3 GHz PXI 5651 85 MHz to 3 3 GHz PXI 5652 85 MHz to 6 6 GHz Store the resulting measurements 8 Compare the measured values to the...

Page 22: ...rough the BNC m to BNC m cable 5 Generate a signal with the following niRFSG property settings Frequency Hz 100 MHz Power Level dBm 0 dBm Generation Mode CW Reference Clock Source ClkIn 6 Use the spec...

Page 23: ...ubharmonic and non integer harmonic levels outside the device frequency range are typical If the results are within the selected test limit the device has passed this portion of the verification Verif...

Page 24: ...uency span 0 Hz Resolution bandwidth 500 Hz Sweep time 5 ms Reference clock source External 7 Use the spectrum analyzer to measure the baseband linearity related spur using the following spectrum anal...

Page 25: ...any available rubidium frequency reference rear panel BNC connector to the spectrum analyzer REF IN rear panel connector through the BNC m to BNC m cable 5 Generate a single sideband tone with a 3 MH...

Page 26: ...limits as specified in Table 13 11 Repeat steps 6 through 10 for the remaining frequencies listed in the Frequency column in Table 13 If the results are within the selected test limit the device has p...

Page 27: ...Verify that the measurement in step 5 is within the 9 ppm test limit with respect to 400 MHz If the results are within the selected test limit the device has passed this portion of the verification Ad...

Page 28: ...Initialize External Calibration VI 2 Call the niRFSG Close External Calibration VI to close the session Set the write calibration to hardware parameter to TRUE to store the results to the EEPROM on th...

Reviews: