NI PXIe-5693 Calibration Procedure
|
© National Instruments
|
45
5.
Connect the NI 5693 EXT FILTER IN connector to the NI 5693 EXT FILTER OUT
connector using the SMA (m)-to-SMA (m) cable. Figure 13 shows the completed
equipment setup.
Figure 13.
Cal Tone Power Accuracy Adjustment Equipment Setup
6.
Create a new session for the NI 5693.
7.
Configure the NI 5693 according to the following settings:
•
External calibration
•
Calibration step: Gain Reference Calibration
•
Preamp: Enabled
•
RF attenuation index: 0
8.
Configure the RF Source 1 according to the following settings:
•
Single frequency mode: Enabled
•
Power level: -25 dBm
9.
Set the power sensor A settling percentage to 0.1% and program the power meter to trigger
when settled.
10. Set the power sensor A as the input sensor and the power sensor B as the output sensor.
11. Turn off the RF source 1 output.
12. Configure the power meter to read only channel B.