RMX-10050/10051 User Guide

The AC output voltages provided by each PDU are based on the mains voltage and the specific 
configuration. AC output ratings are also affected by the specific configuration. 

Note

For the RMX-10051 (785344-01), the neutral pin in the inlet connector is not used.

Table 3.  

RMX-10050/10051 Input and AC Output Ratings

Product

Mains Inlet 

Connector

Mains Inlet 

Rating

AC Output Ratings

RMX-10050
(785341-01)
Single Phase 
with DC Output

IEC 60320 C20

100-240 V~
2P3W 16A
50/60 Hz

C13: 100-240V~, 10 A max per receptacle
Total AC output:
8.7 A max with DC fully loaded, 
16 A max with DC disabled

RMX-10050
(785342-01) 
Single Phase

IEC 60320 C20

100-240 V~
2P3W 16A
50/60 Hz

C13: 100-240V~, 10 A max per receptacle
Total AC output: 16 A max

RMX-10051
(785344-01)
Three Phase 
Wye 4P5W

IEC 60309 
ABL Sursum 
S51S30

220/380-
240/415 V~ L-L
4P5W 16A
50/60 Hz

C13: 220-240 V~, 10 A max per receptacle
C19: 220-240 V~, 16 A max per receptacle
Total AC output
Group J3: 16 A max
Group J2: 16 A max
Group J1: 12.7 A max with DC fully loaded, 
16 A max with DC disabled

RMX-10051
(785343-01)
Three Phase 
Delta 3P4W

IEC 60309 
ABL Sursum
S51S30

200-208 V~ L-L
3P4W 16A 
50/60 Hz

C13: 200-208 V~, 10 A max per receptacle
C19: 200-208 V~, 12.8 A max per receptacle
Total AC output with balanced load
Group J3: 9.2 A max
Group J2: 9.2 A max
Group J1: 5.6 A max with DC fully loaded, 
9.2 A max with DC disabled

Summary of Contents for RMX-10051

Page 1: ...e are two single phase and two three phase PDUs The following tables provide environmental ratings and feature information Table 1 RMX 10050 10051 Environmental Ratings All Models Category Rating Oper...

Page 2: ...A max with DC disabled RMX 10050 785342 01 Single Phase IEC 60320 C20 100 240 V 2P3W 16A 50 60 Hz C13 100 240V 10 A max per receptacle Total AC output 16 A max RMX 10051 785344 01 Three Phase Wye 4P5W...

Page 3: ...plug as well as a redundant chassis ground screw on the rear panel Ensure that the product is properly grounded before applying power Disconnect all power to the product prior to servicing Do no open...

Page 4: ...panel Ensure no external equipment or cables restrict the airflow through the intake or exhaust ports Chassis Ground The rear of the chassis includes a redundant chassis ground screw and ground wire...

Page 5: ...he rear panel The total DC output is always limited to 500 W This is divided evenly as 125 W per each of the four output channels The ratings table in this manual describes the specific outputs and ra...

Page 6: ...ctors use two low voltage signals as follows Shorting the ENABLE pin to the RETURN pin with a dry contact will trigger the Enable signal Opening the DISABLE pin from the RETURN pin will trigger the Di...

Page 7: ...n 44 196 mm 1 74 in Black 27038 All Outside Surfaces Exhaust Air 4 4 X 0 20 DP Holes 2X Both Sides Off Off On On Breaker Breaker Main Breaker Main Breaker GROUP A GROUP A On On Main Power Main Power P...

Page 8: ...d 4 4 X 0 20 DP Holes 3X Both Sides IEC 60309 6H Exhaust Air Main Breaker Main Breaker DC INHIBIT DC INHIBIT J 8 J 8 DC 1 DC 1 DC 2 DC 2 DC 3 DC 3 DC 4 DC 4 J 3 2 J 3 2 J 3 1 J 3 1 J 2 2 J 2 2 J 2 1 J...

Page 9: ...of the European Communities using the manufacturer s declaration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC f...

Page 10: ...at ni com patents You can find information about end user license agreements EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at n...

Reviews: