getting started guide that came with your product for instructions. Remove all cables from the
SCB-100A before getting started.
1.
(Optional) Mount the SCB-100A to a panel or DIN rail, as described in the
2.
Remove the cover.
3.
Remove the film from both sides of the cover.
4.
(Optional) Attach the quick reference label to the inside of the cover. For quick reference
labels for most compatible products, go to
and enter the Info Code
scb100alabels
.
Tip
You can stand the cover in the SCB-100A for easy reference.
5.
Loosen the strain-relief bar by removing the strain-relief screws with a #2 Phillips
screwdriver.
6.
Connect the wires to the screw terminals by stripping off the wire insulation, inserting the
wires into the screw terminals, and securely tightening the screws with the flathead
screwdriver. Refer to the
section for screw-terminal wiring gauge and
torque information.
Caution
To ensure the specified EMC performance, signal wires routed
outside of the enclosure must be contained within a shielded cable and
connected to shielded accessories. Cable shields must be terminated to the
chassis ground lug using as short a connection as is practical.
7.
Reinstall the strain-relief (if removed) and tighten the strain-relief screws. If the shielded
cable is too large to route through the strain-relief hardware, either use multiple,
smaller-diameter cables or remove the top strain-relief bar and add insulation or padding
if necessary to constrain the cable.
8.
Replace the cover.
Caution
You must install cover prior to use.
Caution
Do not connect input voltages >30 V RMS/42 V peak/60 V DC to
the SCB-100A. Input voltages >30 V RMS/42 V peak/60 V DC can damage the
SCB-100A, all devices or modules connected to it, and the host computer.
9.
Connect the SCB-100A to the DAQ device or module using the SH100-100-F cable.
Refer to the
section for information about the SCB-100A I/O connector.
Note
To ensure the specified EMC performance, operate this product only
with shielded cables.
10. Launch Measurement & Automation Explorer (MAX). In the left panel, expand
Devices
and Interfaces
to confirm that your DAQ device or module is recognized, and then
configure your device settings.
SCB-100A User Guide
|
© National Instruments
|