SCXI-1122 Calibration Procedure

DAQchannel

—0

DAQgain

—1

numPoints

—100

Average the resulting scaled data to obtain the final voltage 
measurement.

15. Compare the measurement result to the upper and lower limits shown 

in Table 3. If the measurement value falls between the Upper Limit and 
Lower Limit values, the SCXI-1122 module passed the test.

16. Repeat steps 5 through 15 for the two other test points of the gain 

setting. You only need to check channel 0 on the SCXI-1122 module 
because all inputs are multiplexed back through the same amplifier and 
filter combination.

17. Repeat steps 5 through 16 for each of the remaining gain settings. 

After you have checked all gains for Table 3, continue with step 18.

18. Repeat steps 4 through 17 using Table 3, but set the 

freq

 input 

to 4000.0 in step 7.

Note

If all channels fall within the upper and lower limits, the module does 

not

 need 

adjustment.

You have completed verifying the analog input offsets of the SCXI-1122 
module.

Verifying Excitation Limits

Complete the following steps to verify the excitation limits:

1.

Connect a 15 

Ω

 resistor between VEX+ and VEX– on the front of the 

SCXI module.

Note

If you are not using the SCXI-1322 terminal block, make sure that the SENSE+ 

terminal is connected to the VEX+ terminal and the SENSE– terminal to the VEX– 
terminal.

2.

Set the DMM to voltage mode and connect it across the resistor.

3.

Compare the DMM reading to the upper and lower limits for voltage 
excitation shown in Table 2. If the reading falls between the Upper 
Limit and Lower Limit values, the SCXI-1122 module passed the test.

Summary of Contents for SCXI -1122

Page 1: ...pment and Other Test Requirements 2 Test Equipment 2 Instruments 3 Connectors 3 Software and Documentation 3 Software 3 Documentation 3 Test Conditions 4 Calibration 4 Setting Up the Module 5 Verifyin...

Page 2: ...can affect measurement accuracy as a module ages Calibration restores these components to their specified accuracy and ensures that the module still meets NI standards How Often Should You Calibrate...

Page 3: ...n This section describes the software and documentation you need to calibrate the SCXI 1122 module Software This calibration procedure requires the latest version of the Traditional NI DAQ Legacy driv...

Page 4: ...ent during calibration Keep connections to the SCXI 1122 module and the terminal block short Long cables and wires act as antennae picking up extra noise and thermal offsets that can affect measuremen...

Page 5: ...ct the appropriate calibration interval for your application Verifying Analog Input Offsets Complete the following steps to verify the analog input offsets 1 Read the Test Conditions section in this d...

Page 6: ...e 8 Call SCXI_Single_Chan_Setup to configure the SCXI 1122 module for single channel measurements Set the following parameters SCXIchassisID The value assigned by MAX moduleSlot 1 moduleChan 0 DAQdevi...

Page 7: ...tive and negative inputs of the specified channel For example the positive input for channel 0 is pin A32 which is labeled CH0 The negative input for channel 0 is pin C31 which is labeled CH0 Table 1...

Page 8: ...urned to the buffer variable 14 Call SCXI_Scale to convert the acquired readings into scaled voltage values Set the following parameters SCXIChassisID The value assigned by MAX moduleSlot 0 channel 0...

Page 9: ...gains for Table 3 continue with step 18 18 Repeat steps 4 through 17 using Table 3 but set the freq input to 4000 0 in step 7 Note If all channels fall within the upper and lower limits the module do...

Page 10: ...one for adjusting calibration constants and one for adjusting excitation Adjusting Gain and Offset Errors Complete the following steps to adjust gain and offset errors 1 Ensure that the E Series DAQ d...

Page 11: ...ptable gain values are listed in Table 3 6 Connect the calibrator to analog input channel 0 If you are using an SCXI 1322 terminal block connected to the SCXI 1122 module connect the calibrator to the...

Page 12: ...ator to the negative test point value for the same gain Skip any test point voltages that are specified as 0 V You need only upper and lower limits for adjustment 11 Read the voltage from the DMM Reco...

Page 13: ...llowing steps to adjust the calibration constants 1 Call SCXI_Cal_Constants to create and store the new calibration constants in the memory of the SCXI 1122 module Call this function for each gain set...

Page 14: ...Q device is connected to the SCXI 1122 module 3 Connect a 15 resistor to the VEX and VEX pins which make up the voltage excitation channel Note If you are not using the SCXI 1322 terminal block make s...

Page 15: ...annels of the SCXI 1122 module Verifying Adjusted Values After you complete the adjustment procedure it is important that you verify the accuracy of the adjusted values by repeating the procedure in t...

Page 16: ...1672 48 89579119 49 10420881 0 2000 24 5000 24 55178536 24 44821464 24 5522814 24 4477186 0 0000 0 01434936 0 01434936 0 014845403 0 014845403 24 5000 24 44821464 24 55178536 24 4477186 24 5522814 0 5...

Page 17: ...98121183 0 097868098 0 098131902 100 0000 0 0490 0 049065034 0 048934966 0 049069951 0 048930049 0 0000 3 6712E 05 3 6712E 05 4 16291E 05 4 16291E 05 0 0490 0 048934966 0 049065034 0 048930049 0 04906...

Reviews: