National Instruments Corporation
3
SCXI-1122 Calibration Procedure
Instruments
NI recommends the following instruments for calibrating the SCXI-1122
module:
•
Calibrator—Fluke 5700A
•
DMM—NI 4060 or HP 34401A
If these instruments are not available, use the accuracy requirements listed
above to select substitute calibration instruments.
Connectors
If you do not have custom connection hardware, you need the following
equipment:
•
SCXI-1322 terminal block
•
Shielded 68-pin connector cable
•
50-pin ribbon cable
•
50-pin breakout box
•
SCXI-1349 adapter
These components give easy access to the individual pins on the
SCXI-1122 module front and rear connectors.
Software and Documentation
This section describes the software and documentation you need to
calibrate the SCXI-1122 module.
Software
This calibration procedure requires the latest version of the Traditional
NI-DAQ (Legacy) driver software on the calibration system. The driver
supports a number of programming languages, including LabVIEW and
Measurement Studio. If you do not have the latest version of Traditional
NI-DAQ (Legacy), you can download it from the NI Web site at
ni.com/download
. When you install the driver, you only need to install
support for the programming language used in your application.
Documentation
•
Traditional NI-DAQ Function Reference Help
•
Traditional NI-DAQ User Manual
The
Traditional NI-DAQ Function Reference Help
and the
Traditional
NI-DAQ User Manual
contain detailed information about using Traditional
NI-DAQ (Legacy). The help file includes detailed information about the