If you get the warning message
Failed to find
followed by the module
codes and the message
Unable to communicate with chassis
, take the
following troubleshooting actions:
a.
Make sure that the SCXI chassis is powered on.
b.
Make sure the cable to the SCXI system is properly connected to
a DAQ device.
c.
Inspect the DAQ device connector and SCXI cable adapter for any
bent pins.
d.
Make sure you are using the correct NI cable.
e.
Test the DAQ device to verify it is working properly. See the DAQ
device user manual for more information.
•
If you get the warning message
Failed to find
followed by module
codes and then the message
Instead found: module with ID 0
Xxx
,
return to the
Configuring the SCXI-1141/1142/1143 Module
section
and make sure the correct module is in the specified slot. Delete the
incorrect module as described in the
SCXI-1141/1142/1143 Module from MAX
section and then add the
correct module as described in the
section.
•
If you get the warning message
Failed to find
followed by a module
code and then the message
Slot
x
is empty
, check to see if the
configured module is installed in the specified slot. If not, install the
module by referring to the
Installing the SCXI-1141/1142/1143
section. If the module is installed in the
correct slot, power off the chassis, remove the module as specified in
the
Removing the SCXI-1141/1142/1143 Module from an SCXI
section, and verify that no connector pins are bent on the rear
signal connector. Reinstall the module as shown in the
SCXI-1141/1142/1143 Module in an SCXI Chassis
section, ensuring
that the module is properly aligned in the slot.
•
After checking the preceding items, return to the
section and retest the SCXI system.