Chapter 2
Installing and Configuring the SCXI-1141/1142/1143 Module
SCXI-1141/1142/1143 User Manual
2-6
ni.com
To finish installing the SCXI-1141/1142/1143 module, complete the
following section that is appropriate for your application.
Connecting the SCXI-1141/1142/1143 Module to a DAQ Device for
Multiplexed Scanning in an SCXI Chassis
Use this configuration to multiplex all eight input channels and the
cold-junction compensation (CJC) channel of the SCXI-1141/1142/1143
module into a single channel of the DAQ device. You need the following
items for this installation:
•
SCXI chassis with the SCXI modules installed
•
SCXI cable assembly, which consists of a cable adapter and a cable
•
National Instruments DAQ device installed in the computer
•
1/4 in. flat blade screwdriver
Consult the SCXI chassis documentation, other SCXI module
documentation, and DAQ device documentation for additional instructions
and cautions. The SCXI-1141/1142/1143 module and any other SCXI
modules should already be installed in the chassis according to their
installation instructions.
You can connect the DAQ device to most modules in the chassis that are
configured for multiplexed operation. However, if you use an SCXI-1140,
it must be the cabled module. To connect these modules to the DAQ device
for multiplexed operation, follow these steps while referring to Figure 2-2:
1.
Turn off the SCXI chassis and the computer containing the DAQ
device.
2.
Insert the cable adapter into the back of the SCXI chassis; aligning the
adapter with the module that is to be connected to the DAQ device. See
the installation guide for the cable assembly for more information.
3.
Connect the cable to the back of the cable adapter, ensuring that the
cable fits securely.
4.
Connect the other end of the cable to the DAQ device that you use to
control the SCXI system.
5.
Check the cable installation, making sure the connectors are securely
fastened at both ends.
6.
Turn on the SCXI chassis.
7.
Turn on the computer.
8.
If you have already installed the appropriate software, you are ready to
configure the SCXI-1141/1142/1143 module for multiplexed mode