background image

©

 National Instruments Corporation

7

SCXI-1328 Installation Guide

Installation

To connect the terminal block to the SCXI module front connector, perform 
the following steps:

1.

Connect the module front connector to its mating connector on the 
terminal block.

2.

Tighten the top and bottom thumbscrews on the back of the terminal 
block to hold it securely in place.

Note

For accurate cold-junction compensation, place the SCXI chassis away from an 

extreme temperature differential.

Specifications

Cold-junction sensor

Accuracy

1

........................................ 0.5° from 15 to 35 °C

0.9° from 0° to 15 °C 
and 35 to 55 °C

Repeatability ................................... 0.2° from 15 to 35 °C

Output ............................................. 1.91 to 0.58 V from 0 to 55 °C

Common-mode isolation

Channel to channel.......................... 300 V

rms

Channel to ground........................... 300 V

rms

Environment

Operating temperature............................ 0 to 50 °C

Storage temperature ............................... –20 to 70 °C

Relative humidity ................................... 5% to 90% noncondensing

Safety

Designed in accordance with IEC 61010-1, UL 3111-1, and CAN/CSA 
C22.2 No. 1010.1 for electrical measuring and test equipment. 
Approved at altitudes up to 2000 meters.
Installation Category II
Pollution Degree 2

1   

Includes the combined effects of the temperature sensor accuracy and the temperature difference between the temperature 
sensor and any screw terminal. The temperature sensor accuracy includes tolerances in all component values, the effects 
caused by temperature and loading, and self-heating.

Summary of Contents for SCXI-1328

Page 1: ...put connector The SCXI 1328 has a high precision thermistor and an aluminum isothermal plate to minimize the temperature gradients across the screw terminals when you measure thermocouples The termina...

Page 2: ...so denotes text that is a placeholder for a word or value that you must supply monospace Text in this font denotes text or characters that you should enter from the keyboard sections of code programmi...

Page 3: ...up to 300 V When connecting or disconnecting signal lines to the SCXI terminal block screw terminals make sure the lines are powered off Potential differences between the lines and the SCXI ground cr...

Page 4: ...tance and returns the thermistor temperature Alternatively you can use the following formulas T C TK 273 15 where TK is the temperature in kelvin a 1 295361 10 3 b 2 343159 10 4 c 1 018703 10 7 RT res...

Page 5: ...gnal wires through the strain relief opening You can add insulation or padding if necessary 4 Prepare your signal wire by stripping the insulation no more than 12 mm 5 Connect the wires to the screw t...

Page 6: ...er Lug 4 Mating Connector 5 Thumbscrew 6 Top Cover Screws 7 Top Cover 1 Screw Terminals 2 Berg Connector W1 3 Channel Labeling for the SCXI 1120 D 4 Channel Labeling for the SCXI 1121 2 7 3 6 4 5 1 Ba...

Page 7: ...C Repeatability 0 2 from 15 to 35 C Output 1 91 to 0 58 V from 0 to 55 C Common mode isolation Channel to channel 300 Vrms Channel to ground 300 Vrms Environment Operating temperature 0 to 50 C Stora...

Page 8: ...cost are updated daily and can be found in the Technical Support section of our Web site at www ni com support Worldwide Support National Instruments has offices located around the globe Many branch...

Page 9: ...6 427 Hong Kong 2645 3186 India 91805275406 Israel 03 6120092 Italy 02 413091 Japan 03 5472 2970 Korea 02 596 7456 Mexico D F 5 280 7625 Mexico Monterrey 8 357 7695 Netherlands 0348 433466 Norway 32 2...

Page 10: ......

Reviews: