SCXI-1102/B/C Calibration Procedure
Calibration Process
The calibration process has the following steps:
1.
—Configure the SCXI-1102/B/C for calibration.
2.
—Verify the existing operation of the
SCXI-1102/B/C. This step determines whether the SCXI-1102/B/C is
operating within its test limits.
3.
—Perform an external calibration that adjusts
the SCXI-1102/B/C calibration constants with respect to a known
voltage source.
4.
—Perform another verification to ensure
that the SCXI-1102/B/C is operating within its test limits after
adjustments.
Initial Setup
Complete the following steps to configure the SCXI-1102/B/C for
calibration:
1.
Make sure that all the appropriate driver and application software is
installed.
2.
Make sure all components involved in the calibration procedure are
powered off.
3.
Assemble the SCXI-1102/B/C, SCXI chassis, terminal blocks, and
DAQ device as shown in Figure 1. You must cable the SCXI-1102/B/C
directly to the DAQ device.